CLASS I Device recall · Reported 4 Oct 2017 · FDA Enforcement Report

Bridge Occlusion Balloon catheter, Sheath Compatibility 12F recalled

Spectranetics is recalling Bridge Occlusion Balloon catheter, Sheath Compatibility 12F, Guidewire Compatibility, distributed nationwide in the US. The recall was initiated on 7 Aug 2017 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesModel Number 590-001, Lots FMN17B13A, FMN17C08A, FMN17C28A, FMN17D07A, FMN47D12A, FMN17D19A, FMN17D27A, FMN17E02A, FMN17E23A, FMN17E31A, FMN17E31B, FMN17F06A, FMN17F20A, FMN17F21A, FMN17G12A, FMN17G18A. Additional lots: FMN17H03A, FMN17H04A, FMN17H05A, FMN17H10A, FMN17H12A, FMN17H15A, FMN17H17A, FMN17H29A, FMN17J28A, FMN17K02A.
Quantity recalled2,356 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3106-2017
FDA event ID
77934
Recalling firm
Spectranetics, Colorado Springs, CO
Classification
Class I
Status
Terminated
Recall initiated
7 Aug 2017
FDA classified
22 Sep 2017
Reported
4 Oct 2017
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

7 Aug 2017Recall initiated by company
22 Sep 2017Classified by FDA+46 days
4 Oct 2017Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Possible inability to pass the guidewire through the Bridge device, therefore, resulting in a potential delay of treatment.

Product as listed by the FDA

Spectranetics Bridge Occlusion Balloon catheter, Sheath Compatibility 12F, Guidewire Compatibility 0.035in, Working Length 90cm, Balloon Outer Diameter 31mm, Balloon Length 80mm, Rx Only. Product Usage: It is a device used to occlude the vessel and prevent blood loss as the physician preps for surgery resulting from an SVC tear complication of a lead extraction case. It is used for temporary vessel occlusion of the superior vena cava in applications including perioperative occlusion and emergency control of hemorrhage.

Where it was distributed

Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Cyprus, Denmark, Finland, France, Germany, Hungary, Iceland, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Poland, South Africa, Sweden, Switzerland, and United Kingdom.

Nationwide

Questions about this recall

Why was it recalled?

Possible inability to pass the guidewire through the Bridge device, therefore, resulting in a potential delay of treatment.

Which lots are affected?

Model Number 590-001, Lots FMN17B13A, FMN17C08A, FMN17C28A, FMN17D07A, FMN47D12A, FMN17D19A, FMN17D27A, FMN17E02A, FMN17E23A, FMN17E31A, FMN17E31B, FMN17F06A, FMN17F20A, FMN17F21A, FMN17G12A, FMN17G18A. Additional lots: FMN17H03A, FMN17H04A, FMN17H05A, FMN17H10A, FMN17H12A, FMN17H15A, FMN17H17A, FMN17H29A, FMN17J28A, FMN17K02A.

Where was it sold?

Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Cyprus, Denmark, Finland, France, Germany, Hungary, Iceland, Ireland, Israel, Italy, Kuwait, Netherlands, Norway, Poland, South Africa, Sweden, Switzerland, and United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.