CLASS II ONGOING Device recall · Reported 13 Feb 2019 · FDA Enforcement Report

Bronchoscope Tubes recalled by Richard Wolf

Richard Wolf is recalling Bronchoscope Tubes, distributed nationwide in the US. The recall was initiated on 17 Oct 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll lot numbers manufactured from October 30, 2006 to August 15, 2018, with affected Product and Lot Numbers distributed worldwide - Product Number (Lot Numbers): 825200641 (1271924, 1074548, 1186516, 1364805, 1262505, 1269165, 1313148, 1296901, 1330427, 1382941, 1321674, 1339578, 1265501, 1344236, 1261415, 1372437, 1353961, M054680, M118520, 1003492, 1058979, 1287750); 825200842 (1189516, 1074551, 1319855, 1298913, 1057370, M054700, M118540); 825201041 (1269824, 1277860, 1186520, 1195261, 1230705, 1340401, 1253661, 1263078, 1288520, 1186520 , 1227431, 1368550, 1376018, 1382289, 1316167, 1104163, 1074553, 1307329, 1296285, 1332473, 1319856, 1053400, 1337541, M054720, M118550); 825201241 (1185831, 1372500, 1364881, 1238365, 1273017, 1302457, 1298771, 1186521, 1269168, 1183846, 1323532, 1074555, 1316169, 1344797, 1333774, 1057371, 1361470, 1265502, 1289744, M054730, M118560); 825201441 (1185832, 1189518, 1339287, 1274102, 1281289, 1320136, 1314041, 1360269, 1328548, 1003493, 1074556, M054740, M112910); 825210830 (1264551, 1313443, 1338468, 1212278, 1262656, 1271357, 1320727); 825211030 (1271954, 1293167, 1287813, 1313444, 1265025, 1320728, 1212279, 1342362); 825211230 (1240940, 1253421, 1245260); 825211430 (1253422, 1240941); 825200841 (1274105, 1185829, 1281667, 1257138, 1368638, 1306948, 1265989, 1296284, 1298436, 1189515, 1269820, 1319630, 1376097, 1362251, 1323968, 1107980, 1333771, 1261416, 1378166, 1053398, 1338467, 1288856, 1074550, M054690, M118530, 1241165, 1243611, 1248730, 1253658, 1262506); Affected Part and Lot numbers distributed Nationwide - Part Number (Lot Numbers): 82520.0641 (1058979, 1074548, 1339578, 1003492, 1313148, M054680, 1186516, 1271924, 1262505, 1269165, 1296901, 1364805, 1330427, 1344236, M118520, 1287750, 1353961, 1382941); 82520.0841 (1074550, 1333771, 1323968, M054690, 1281667, 1274105, 1241165, 1053398, 1269820, 1243611, 1248730, 1262506, 1288856, 1298436, M118530, 1362251, 1185829, 1198515, 1338467, 1378166, 1253658, 1296284); 82520.0842 (M118540, M054700, 1189516, 1319855); 82520.1041 (1074553, 1319856, 1316167, 1277860, M054720, 1053400, 1269824, 1230705, 1253661, 1263078, 1376018, 1288520, 1332473, M118550, 1337541, 1186520, 1195261, 1382289); 82520.1241 (1074555, 1333774, 1302457, 1057371, M054730, 1269168, M118560, 1238365, 1186521, 1289744, 1372500, 1298771, 1344797, 1361470, 1265502, 1185831); 82520.1441 (1074556, 1320136, 1003493, 1314041, 1281289, M054740, M112910, 1189518, 1360269, 1274102, 1339287, 1328548); 825210830 (1271357, 1313443, 1212278, 1262656, 1320727, 1264551, 1338468); 825211030 (1320728, 1313444, 1212279, 1265025, 1342362, 1293167, 1287813); 825211230 (1245260, 1253421, 1240940); 825211430 (1240941, 1253422)
Quantity recalled873 units worldwide (468 units within the US)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0804-2019
FDA event ID
81169
Recalling firm
Richard Wolf, Knittlingen
Classification
Class II
Status
Ongoing
Recall initiated
17 Oct 2018
FDA classified
6 Feb 2019
Reported
13 Feb 2019
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

17 Oct 2018Recall initiated by company
6 Feb 2019Classified by FDA+112 days
13 Feb 2019Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential deficiency of the TEXAS Bronchoscope Tube product. The non-functionally relevant bonding points / cemented areas of the Bronchoscope Tubes may loosen.

Product as listed by the FDA

Richard Wolf Bronchoscope Tubes, Model Texas. Device is reusable and should be sterilized prior to each use. Product Usage: Bronchoscopes are intended for investigation, diagnosis and therapy of pathological changes and removal of foreign bodies in the area of the tracheobronchial respiratory pathways, in conjunction with endoscopic accessories. The bronchoscopes can also be used endoscopically to visualize the tracheobionchial respiratory tract via the natural passage (oral). Tracheoscope tubes are used as follows: 1. The tube is used as a passage to the area to be examined or treated and allows the insertion of auxiliary instruments such as forceps, resector, remover , suction device and stents; 2. The internal diameter allows the conveying of gases for artificial respiration of the patient and via the ventilation connector in conjunction with the telescope element, the tube can be connected to all known respiration and anasthetic systems; 3. Sampling channel for respiratory gas analysis when gas for analysis is gained from the patient s central respiratory tract by means of the bypass method via the tracheoscope tube. M = identification on Luer connector; 4.Tracheoscopes are used in the trachea up to the tracheal bifurcation.

Where it was distributed

Worldwide distribution - US Nationwide in the states of PA, MD, NY, and countries of Bangladesh, Canada, China, France, Germany, India, Italy, Jordan, Netherlands, Poland, Russia, Spain, Thailand, Turkey.

Nationwide MarylandNew YorkPennsylvania

Questions about this recall

Why was it recalled?

There is a potential deficiency of the TEXAS Bronchoscope Tube product. The non-functionally relevant bonding points / cemented areas of the Bronchoscope Tubes may loosen.

Which lots are affected?

All lot numbers manufactured from October 30, 2006 to August 15, 2018, with affected Product and Lot Numbers distributed worldwide - Product Number (Lot Numbers): 825200641 (1271924, 1074548, 1186516, 1364805, 1262505, 1269165, 1313148, 1296901, 1330427, 1382941, 1321674, 1339578, 1265501, 1344236, 1261415, 1372437, 1353961, M054680, M118520, 1003492, 1058979, 1287750); 825200842 (1189516, 1074551, 1319855, 1298913, 1057370, M054700, M118540); 825201041 (1269824, 1277860, 1186520, 1195261, 1230705, 1340401, 1253661, 1263078, 1288520, 1186520 , 1227431, 1368550, 1376018, 1382289, 1316167, 1104163, 1074553, 1307329, 1296285, 1332473, 1319856, 1053400, 1337541, M054720, M118550); 825201241 (1185831, 1372500, 1364881, 1238365, 1273017, 1302457, 1298771, 1186521, 1269168, 1183846, 1323532, 1074555, 1316169, 1344797, 1333774, 1057371, 1361470, 1265502, 1289744, M054730, M118560); 825201441 (1185832, 1189518, 1339287, 1274102, 1281289, 1320136, 1314041, 1360269, 1328548, 1003493, 1074556, M054740, M112910); 825210830 (1264551, 1313443, 1338468, 1212278, 1262656, 1271357, 1320727); 825211030 (1271954, 1293167, 1287813, 1313444, 1265025, 1320728, 1212279, 1342362); 825211230 (1240940, 1253421, 1245260); 825211430 (1253422, 1240941); 825200841 (1274105, 1185829, 1281667, 1257138, 1368638, 1306948, 1265989, 1296284, 1298436, 1189515, 1269820, 1319630, 1376097, 1362251, 1323968, 1107980, 1333771, 1261416, 1378166, 1053398, 1338467, 1288856, 1074550, M054690, M118530, 1241165, 1243611, 1248730, 1253658, 1262506); Affected Part and Lot numbers distributed Nationwide - Part Number (Lot Numbers): 82520.0641 (1058979, 1074548, 1339578, 1003492, 1313148, M054680, 1186516, 1271924, 1262505, 1269165, 1296901, 1364805, 1330427, 1344236, M118520, 1287750, 1353961, 1382941); 82520.0841 (1074550, 1333771, 1323968, M054690, 1281667, 1274105, 1241165, 1053398, 1269820, 1243611, 1248730, 1262506, 1288856, 1298436, M118530, 1362251, 1185829, 1198515, 1338467, 1378166, 1253658, 1296284); 82520.0842 (M118540, M054700, 1189516, 1319855); 82520.1041 (1074553, 1319856, 1316167, 1277860, M054720, 1053400, 1269824, 1230705, 1253661, 1263078, 1376018, 1288520, 1332473, M118550, 1337541, 1186520, 1195261, 1382289); 82520.1241 (1074555, 1333774, 1302457, 1057371, M054730, 1269168, M118560, 1238365, 1186521, 1289744, 1372500, 1298771, 1344797, 1361470, 1265502, 1185831); 82520.1441 (1074556, 1320136, 1003493, 1314041, 1281289, M054740, M112910, 1189518, 1360269, 1274102, 1339287, 1328548); 825210830 (1271357, 1313443, 1212278, 1262656, 1320727, 1264551, 1338468); 825211030 (1320728, 1313444, 1212279, 1265025, 1342362, 1293167, 1287813); 825211230 (1245260, 1253421, 1240940); 825211430 (1240941, 1253422)

Where was it sold?

Worldwide distribution - US Nationwide in the states of PA, MD, NY, and countries of Bangladesh, Canada, China, France, Germany, India, Italy, Jordan, Netherlands, Poland, Russia, Spain, Thailand, Turkey.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.