CLASS II ONGOING Device recall · Reported 13 Apr 2022 · FDA Enforcement Report
Cardiohelp-i Transport Guard with Transport Guard recalled
Maquet Medical Systems USA is recalling Cardiohelp-i Transport Guard with Transport Guard (701072525): is a blood oxygenation and, distributed nationwide in the US. The recall was initiated on 18 Feb 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0897-2022
- FDA event ID
- 89745
- Recalling firm
- Maquet Medical Systems USA, Wayne, NJ
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 18 Feb 2022
- FDA classified
- 8 Apr 2022
- Reported
- 13 Apr 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 18 Feb 2022 | Recall initiated by company | |
| 8 Apr 2022 | Classified by FDA | +49 days |
| 13 Apr 2022 | Published in enforcement report | +5 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Potential for a loose cylindrical pin on the Transport Guard that may result in the jamming of the locking mechanism causing the Transport Guard to not open or, if the locking pin is fully disengaged from the assembly, a full disassembly of the Transport Guard may occur, resulting in a delay of support if a product exchange is necessary.
Product as listed by the FDA
Cardiohelp-i Transport Guard with Transport Guard (701072525): is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support for periods appropriate to cardiopulmonary bypass (up to six hours). Product Code/REF Number: 701072780 (Cardiohelp-i)
Where it was distributed
US Nationwide Distribution: FL, KY, NH, NJ, NY, PA, TX, VA Foreign: Belgium, Canada, Germany, Hungary, Luxemburg, Malaysia, Mexico, Poland, Slovakia, South Africa, South Korea, and Taiwan.
Nationwide FloridaKentuckyNew HampshireNew JerseyNew YorkPennsylvaniaTexasVirginia
Questions about this recall
Why was it recalled?
Potential for a loose cylindrical pin on the Transport Guard that may result in the jamming of the locking mechanism causing the Transport Guard to not open or, if the locking pin is fully disengaged from the assembly, a full disassembly of the Transport Guard may occur, resulting in a delay of support if a product exchange is necessary.
Which lots are affected?
Serial Numbers for the Cardiohelp-i (Part Number 701072780)as the Transport Guard (701072525) is packaged with the Cardiohelp-i and is not labelled individually Serial Numbers: 90413945, 90414225, 90414226, 90414227, 90414228, 90414229, 90414232, 90414233, 90414234, 90414235, 90414236, 90414237, 90414283, 90414284, 90414285, 90414286, 90414287, 90414288, 90414289, 90414292, 90414293, 90414294, 90414295, 90414296, 90414297, 90414309, 90414310, 90414311, 90414312, 90414317, 90414318, 90414319. UDI: 04058863074863 (Cardiohelp-i)
Where was it sold?
US Nationwide Distribution: FL, KY, NH, NJ, NY, PA, TX, VA Foreign: Belgium, Canada, Germany, Hungary, Luxemburg, Malaysia, Mexico, Poland, Slovakia, South Africa, South Korea, and Taiwan.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Maquet Medical Systems USA
All 41| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Cardiohelp-i System recalled by Maquet Medical Systems U | Maquet Medical Systems USA | Packaging Defects | Nationwide |
| Device | CLASS II | Model No. Beq-hls 5050 USA, HLS Set Advanced 5.0, Product Code/Part recalled | Maquet Medical Systems USA | Sterility | Nationwide |
| Device | CLASS II | Beq-hls 7050 USA, HLS Set Advanced 7.0, Product recalled | Maquet Medical Systems USA | Sterility | Nationwide |
| Device | CLASS II | Beq-rf-32-Usa RotaFlow Centrifugal Pump recalled | Maquet Medical Systems USA | Sterility | Nationwide |
| Device | CLASS II | Bo-rf-32-Usa RotaFlow Centrifugal Pump recalled | Maquet Medical Systems USA | Sterility | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |