CLASS II Device recall · Reported 12 Aug 2020 · FDA Enforcement Report
Carescape ONE, Physiological Patient Monitor recalled by GE Healthcare
GE Healthcare is recalling Carescape ONE, Physiological Patient Monitor, distributed nationwide in the US. The recall was initiated on 22 Jun 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2750-2020
- FDA event ID
- 85979
- Recalling firm
- GE Healthcare, Waukesha, WI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 22 Jun 2020
- FDA classified
- 4 Aug 2020
- Reported
- 12 Aug 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Patient Monitors
Timeline
| 22 Jun 2020 | Recall initiated by company | |
| 4 Aug 2020 | Classified by FDA | +43 days |
| 12 Aug 2020 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
CARESCAPE ONE may not provide visual and audible alarms for Ventricular Fibrillation (V Fib), if V Fib occurs at the time CARESCAPE ONE is docked to a CARESCAPE B450/B650/B850 host monitor.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
CARESCAPE ONE, Physiological Patient Monitor - Product Usage: is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility.
Where it was distributed
Worldwide distribution - US Nationwide distribution including in the states of MT, NC, TX and WI. The countries of Australia, Austria, Bahrain,Belgium, Bolivia, Brazil ,Canada, Chile, Check Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Malta, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Russia, Saudi Arabia, Singapore, Saint Martin, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirates, United Kingdom.
Questions about this recall
Is Carescape ONE, Physiological Patient Monitor recalled?
Yes. GE Healthcare recalled Carescape ONE, Physiological Patient Monitor on 22 Jun 2020. The recall is Class II and its status is Terminated. Recall number Z-2750-2020.
Why was it recalled?
CARESCAPE ONE may not provide visual and audible alarms for Ventricular Fibrillation (V Fib), if V Fib occurs at the time CARESCAPE ONE is docked to a CARESCAPE B450/B650/B850 host monitor.
Which lots are affected?
Model: 2087075-001 Serial Numbers: SNA20210164SA, SNA20210031SA, SNA20210045SA, SNA20210064SA, SNA20210319SA, SNA20210330SA, SNA20210348SA, SNA20210361SA, SNA20210362SA, SNA20210375SA, SNA20210376SA, SNA20210377SA, SNA20220003SA, SNA20220005SA, SNA20220007SA, SNA20220018SA, SNA20220022SA, SNA20220026SA, SNA20220045SA, SNA20220047SA, SNA18140024SA, SNA18340072SA, SNA18370025SA, SNA18380001SA, SNA18390009SA, SNA18430002SA, SNA18430059SA, SNA18430088SA, SNA18430116SA, SNA19390023SA, SNA19400056SA, SNA19400143SA, SNA19440048SA, SNA19440069SA, SNA19440071SA, SNA19440122SA, SNA19450008SA, SNA19450025SA, SNA19450128SA, SNA19450136SA, SNA19500072SA, SNA19500073SA, SNA19500080SA, SNA19500084SA, SNA19500086SA, SNA19510001SA, SNA19510005SA, SNA19510007SA, SNA19510012SA, SNA19510027SA, SNA19510053SA, SNA19510055SA, SNA19510103SA, SNA20020119SA, SNA20030070SA, SNA20030071SA, SNA20170022SA, SNA20170080SA, SNA20170082SA, SNA20170112SA, SNA20170113SA, SNA20170116SA, SNA20170117SA, SNA20170118SA, SNA20170120SA, SNA20170131SA, SNA20170134SA, SNA20170136SA, SNA20170139SA, SNA20170145SA, SNA20170160SA, SNA20170161SA, SNA20170164SA, SNA20170166SA, SNA20170168SA, SNA20170169SA, SNA20170171SA, SNA20170173SA, SNA20170176SA, SNA20170181SA, SNA20170182SA, SNA20170194SA, SNA20170197SA, SNA20170198SA, SNA20170232SA, SNA20170237SA, SNA20170264SA, SNA20180002SA, SNA20180008SA, SNA20180009SA, SNA20180011SA, SNA20180012SA, SNA20180017SA, SNA20180018SA, SNA20180023SA, SNA20180025SA, SNA20180026SA, SNA20180027SA, SNA20180028SA, SNA20180029SA, SNA20180030SA, SNA20180031SA, SNA20180032SA, SNA20180040SA, SNA20180042SA, SNA20180043SA, SNA20180044SA, SNA20180071SA, SNA20180073SA, SNA20180074SA, SNA20180075SA, SNA20180076SA, SNA20180080SA, SNA20180096SA, SNA20180101SA, SNA20180102SA, SNA20180109SA, SNA20180117SA, SNA20180144SA, SNA20180161SA, SNA20180162SA, SNA20190321SA, SNA20190331SA, SNA18140004SA, SNA18140005SA, SNA18140007SA, SNA18240025SA, SNA18240026SA, SNA18240030SA, SNA18240032SA, SNA18240044SA, SNA18240051SA, SNA18240054SA, SNA18240079SA, SNA18250010SA, SNA18250012SA, SNA18250017SA, SNA18250023SA, SNA18250026SA, SNA18250033SA, SNA18250034SA, SNA18250036SA, SNA18250044SA, SNA18250061SA, SNA18250072SA, SNA18250075SA, SNA18250078SA, SNA18250080SA, SNA18250081SA, SNA18250082SA, SNA18250086SA, SNA18250092SA, SNA18250093SA, SNA18440051SA, SNA18450094SA, SNA18450098SA, SNA18460013SA, SNA18460040SA, SNA18460072SA, SNA18460075SA, SNA18460083SA, SNA18460086SA, SNA18460089SA, SNA18460101SA, SNA19040008SA, SNA19050016SA, SNA19050025SA, SNA19050028SA, SNA19050030SA, SNA19050031SA, SNA19050033SA, SNA19050038SA, SNA19050043SA, SNA19050044SA, SNA19050045SA, SNA19050046SA, SNA19050049SA, SNA19050050SA, SNA19070071SA, SNA19070072SA, SNA19070073SA, SNA19070074SA, SNA19490074SA, SNA19500037SA, SNA19500043SA, SNA19500054SA, SNA19500059SA, SNA19510003SA, SNA20140029SA, SNA20140031SA, SNA20190324SA, SNA19500131SA, SNA20030120SA, SNA20040052SA, SNA20170179SA, SNA20170195SA, SNA20170203SA, SNA20180054SA, SNA20180057SA, SNA20180059SA, SNA20180077SA, SNA20180100SA, SNA20180103SA, SNA20180104SA, SNA20180113SA, SNA20180115SA, SNA20180142SA, SNA20180145SA, SNA20180146SA, SNA20180147SA, SNA20180154SA, SNA20180158SA, SNA20180163SA, SNA20180168SA, SNA20180183SA, SNA20180202SA, SNA20190024SA, SNA20190026SA, SNA20190030SA, SNA20190034SA, SNA20190038SA, SNA20190056SA, SNA20190057SA, SNA20190059SA, SNA20190060SA, SNA20190063SA, SNA20190106SA, SNA20190107SA, SNA20190109SA, SNA20190112SA, SNA20190114SA, SNA20190115SA, SNA20190116SA, SNA20190117SA, SNA20190119SA, SNA20190120SA, SNA20190121SA, SNA20190122SA, SNA20190124SA, SNA20190126SA, SNA20190127SA, SNA20190130SA, SNA20190131SA, SNA20190132SA, SNA20190133SA, SNA20190134SA, SNA20190137SA, SNA20190141SA, SNA20190142SA, SNA20190143SA, SNA20190144SA, SNA20190145SA, SNA20190146SA, SNA20190147SA, SNA20190148SA, SNA20190149SA, SNA20190152SA, SNA20190153SA, SNA20190155SA, SNA20190156SA, SNA20190159SA, SNA20190161SA, SNA20190174SA, SNA20190201SA, SNA20190204SA, SNA20190206SA, SNA20190209SA, SNA20190210SA, SNA20190214SA, SNA20190216SA, SNA20190217SA, SNA20190218SA, SNA20190219SA, SNA20190220SA, SNA20190221SA, SNA20190222SA, SNA20190223SA, SNA20190229SA, SNA20190230SA, SNA20190231SA, SNA20190232SA, SNA20190233SA, SNA20190235SA, SNA20190236SA, SNA20190237SA, SNA20190238SA, SNA20190239SA, SNA20190240SA, SNA20190247SA, SNA20190279SA, SNA20190289SA, SNA20190328SA, SNA20190339SA, SNA20200031SA, SNA20200081SA, SNA20200403SA.
Where was it sold?
Worldwide distribution - US Nationwide distribution including in the states of MT, NC, TX and WI. The countries of Australia, Austria, Bahrain,Belgium, Bolivia, Brazil ,Canada, Chile, Check Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kenya, Korea, Kuwait, Latvia, Lebanon, Lithuania, Macedonia, Malaysia, Malta, Morocco, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Russia, Saudi Arabia, Singapore, Saint Martin, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Arab Emirates, United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from GE Healthcare
All 512| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Signa Premier systems recalled by GE Healthcare | GE Healthcare | Other | Nationwide |
| Device | CLASS II | Centricity Universal Viewer Zero Footprint Client recalled | GE Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | GE HealthCare ViewPoint 6, Catalog Number H47601AA; Radiological recalled | GE Healthcare | Device Malfunction | |
| Device | CLASS II | Centricity Pacs-iw (Pacs-iw) recalled by GE Healthcare | GE Healthcare | Other | IL |
| Device | CLASS II | Centricity Centricity Pacs-iw with Universal Viewer recalled | GE Healthcare | Other | IL |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |