CLASS II ONGOING Device recall · Reported 2 Sep 2026 · FDA Enforcement Report

IQecg Digital ECG recalled by Midmark

Midmark is recalling IQecg Digital ECG, distributed nationwide in the US. The recall was initiated on 20 Jul 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark IQecg Digital ECG. The Midmark IQecg Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 480000 4820704 UDI-DI: 00841709100581, 00841709100321, 00841709100895, 00841709100901
Quantity recalled3624 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3012-2026
FDA event ID
99404
Recalling firm
Midmark, Versailles, OH
Classification
Class II
Status
Ongoing
Recall initiated
20 Jul 2026
FDA classified
25 Aug 2026
Reported
2 Sep 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors
Device type
Patient Monitors

Timeline

20 Jul 2026Recall initiated by company
25 Aug 2026Classified by FDA+36 days
2 Sep 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.

Product as listed by the FDA

Midmark IQecg Digital ECG, Model/Catalog Number: 4-000-0061, 4-000-0062 Software Version: Midmark IQecg Plugin versions starting with v10 and prior to v10.1.0.18.

Where it was distributed

Worldwide distribution - US Nationwide and the country of Canada.

Nationwide

Questions about this recall

Why was it recalled?

A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.

Which lots are affected?

Lot Code: This correction affects IQecg plugin software versions starting with v10 and prior to v10.1.0.18 installed on Midmark IQecg Digital ECG. The Midmark IQecg Digital ECG system information is provided below. The total quantities manufactured and distributed noted below are the combined IQecgs Digital ECG and Midmark Digital ECG systems. Serial Numbers: 480000 4820704 UDI-DI: 00841709100581, 00841709100321, 00841709100895, 00841709100901

Where was it sold?

Worldwide distribution - US Nationwide and the country of Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 99404

This product is part of a recall event; the main page for the event is Digital ECG HD+12, HD+15 recalled by Midmark.

ReportedClassRecallCompanyReasonWhere
CLASS IIDigital ECG HD+12, HD+15 recalled by MidmarkMidmarkLabeling ErrorsNationwide

More recalls from Midmark

All 8

Other Patient Monitors recalls

All 1,014