CLASS II ONGOING Device recall · Reported 21 Jan 2026 · FDA Enforcement Report
Catalyst, Catalyst+ recalled by C-rad Positioning Ab
C-rad Positioning AB is recalling Catalyst, Catalyst+, distributed nationwide in the US. The recall was initiated on 9 Oct 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1092-2026
- FDA event ID
- 97872
- Recalling firm
- C-rad Positioning AB, Uppsala
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 9 Oct 2025
- FDA classified
- 14 Jan 2026
- Reported
- 21 Jan 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Imaging Equipment
Timeline
| 9 Oct 2025 | Recall initiated by company | |
| 14 Jan 2026 | Classified by FDA | +97 days |
| 21 Jan 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Brand Name: Catalyst+ Product Name: Catalyst, Catalyst+ Model/Catalog Number: SP001-0026, SP003-0002 Software Version: cSRS module in combination with cMotion module Product Description: GMDN: 40893 Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not applicable
Where it was distributed
The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.
Nationwide AlaskaAlabamaArkansasCaliforniaConnecticutFloridaMissouriNorth CarolinaNew MexicoNew YorkOhioOklahomaOregonPennsylvaniaTennesseeTexasVirginiaWashington
Questions about this recall
Why was it recalled?
Due to issues with the system's stereotactic radiosurgery (SRS) treatment cannot be guaranteed for all couch angles and system setups. n certain cases, the system may indicate the patient as correctly positioned at the isocenter, even when the isocenter is positioned outside of the indicated tolerance.
Which lots are affected?
Lot Code: Note: The error is not related to any serial numbers or LOT numbers. The error is limited to the software module cSRS in combination with cMotion module
Where was it sold?
The software module cSRS has been installed in 37 countries Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, CA, CT, FL, MO, NC, NM, NY, OH, OK, OR, PA, TN, TX, VA, and WA. The countries of Argentina, Australia, Austria, Belarus, Belgium, Benin, Canada, China, Colombia, Egypt, El Salvador, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Jordan, Mexico, Morocco, Netherlands, New Zealand, Norway, Portugal, Romania, Russia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Ukraine, and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from C-rad Positioning AB
All 3| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Brand Name: Catalyst+ Product Name: C4D software used in conjunction recalled | C-rad Positioning AB | Device Malfunction | Nationwide |
| Device | CLASS II | PC Application recalled by C-rad Positioning Ab | C-rad Positioning AB | Device Malfunction | Nationwide |
Other Imaging Equipment recalls
All 2,347| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |