CLASS II Device recall · Reported 14 Sep 2016 · FDA Enforcement Report

Centricity Enterprise Archive recalled by Ge Healthcare It

Ge Healthcare It is recalling Centricity Enterprise Archive is a, distributed nationwide in the US. The recall was initiated on 16 Nov 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesVersion: 3.0.x
Quantity recalled1,379 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2739-2016
FDA event ID
64761
Recalling firm
Ge Healthcare It, Barrington, IL
Classification
Class II
Status
Terminated
Recall initiated
16 Nov 2012
FDA classified
6 Sep 2016
Reported
14 Sep 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

16 Nov 2012Recall initiated by company
6 Sep 2016Classified by FDA+1390 days
14 Sep 2016Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There is a potential for data loss associated with empty path names resulting from HL-7 updates when the cache disk is almost full. In an unlikely event that the cache disk is almost full (<100MB) and HL-7 update events are received, the update action fails and an empty path name is written to the EA database. As a result the instance is not deleted from the storage device but no longer referenced. When trying to retrieve the instance an error is returned.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Centricity Enterprise Archive is a software product for receiving, archiving and sending of medical data. Qualified system administrators install, monitor and maintain the system. DICOM devices (e.g. modalities, workstations) communicate with the archive using the DICOM protocol (published by ACR-NEMA). The clinical use is limited to the interaction other systems have with the Centricity Enterprise Archive. Centricity Enterprise Archive is not intended for use in patient vicinity. The intended users for Centricity Enterprise Users service personnel and System administrators, and non clinical users. As of V4.0 XDS enabled systems can communicate with the archive using the XDS and XDS-I profiles (published by IHE).

Where it was distributed

Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

There is a potential for data loss associated with empty path names resulting from HL-7 updates when the cache disk is almost full. In an unlikely event that the cache disk is almost full (<100MB) and HL-7 update events are received, the update action fails and an empty path name is written to the EA database. As a result the instance is not deleted from the storage device but no longer referenced. When trying to retrieve the instance an error is returned.

Which lots are affected?

Version: 3.0.x

Where was it sold?

Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Ge Healthcare It

All 17

Other Patient Monitors recalls

All 1,022