CLASS II Device recall · Reported 24 Aug 2016 · FDA Enforcement Report

GE Centricity PACS Workstation recalled by GE Healthcare It

GE Healthcare It is recalling GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and, distributed nationwide in the US. The recall was initiated on 15 Nov 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCentricity PACS versions 3.1.1.x through 3.2.1.x
Quantity recalled498 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2579-2016
FDA event ID
64785
Recalling firm
GE Healthcare It, Barrington, IL
Classification
Class II
Status
Terminated
Recall initiated
15 Nov 2012
FDA classified
17 Aug 2016
Reported
24 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

15 Nov 2012Recall initiated by company
17 Aug 2016Classified by FDA+1371 days
24 Aug 2016Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

While merging exams in a test instance, merges carry forward in the production system, when the middle tier is configured to the wrong IP address of the Centricity Exam Manager. When a current and historical exam are opened at the same time, the system asynchronously tries to access the operation specifying how each image should be grouped, causing random image-series grouping errors.

Product as listed by the FDA

GE Centricity PACS Workstation, versions 3.1.1.x through 3.2.1.x Picture Archiving and Communication System Used as a primary diagnostic and analysis tool for diagnostic images by trained healthcare professionals, including radiologists, physicians, technologists, clinicians and nurses. Also used as a clinical review workstation throughout the healthcare facility. The workstation interface provides the user with a means to display, manipulate, archive, print and export images when connected with the Centricity PACS infrastructure.

Where it was distributed

Worldwide distribution including US nationwide, Puerto Rico, Australia, Austria, Belgium, Bermuda, CANADA, Denmark, Egypt, ENGLAND, France, Germany, India, Ireland, Hong Kong, Korea, Kuwait, Italy, Malta, Netherlands, Portugal, Qatar, Saudi Arabia, SCOTLAND, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, and UK.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

While merging exams in a test instance, merges carry forward in the production system, when the middle tier is configured to the wrong IP address of the Centricity Exam Manager. When a current and historical exam are opened at the same time, the system asynchronously tries to access the operation specifying how each image should be grouped, causing random image-series grouping errors.

Which lots are affected?

Centricity PACS versions 3.1.1.x through 3.2.1.x

Where was it sold?

Worldwide distribution including US nationwide, Puerto Rico, Australia, Austria, Belgium, Bermuda, CANADA, Denmark, Egypt, ENGLAND, France, Germany, India, Ireland, Hong Kong, Korea, Kuwait, Italy, Malta, Netherlands, Portugal, Qatar, Saudi Arabia, SCOTLAND, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, and UK.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.