CLASS II ONGOING Device recall · Reported 13 Nov 2024 · FDA Enforcement Report

Centricity Universal Viewer Zero Footprint recalled by GE Healthcare

GE Healthcare is recalling Centricity Universal Viewer Zero Footprint, distributed nationwide in the US. The recall was initiated on 3 Oct 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI/DI 00840682102988 for all models: 1) 2089507-083, Installed Product ID Numbers: GDZH-JINWAN-Centricity Universal Viewer Zero Footprint / ZFP-02171477, 103378-Centricity Universal Viewer Zero Footprint / ZFP-20039058, 1664823-Centricity Universal Viewer Zero Footprint / ZFP-20080654, 493356-Centricity Universal Viewer Zero Footprint / ZFP-20085691, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291850, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291877, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02292580, 81077-Centricity Universal Viewer Zero Footprint / ZFP-20680055, IL1469-Centricity Universal Viewer Zero Footprint / ZFP-02156638, JP-110145-Centricity Universal Viewer Zero Footprint / ZFP-01275393, 1333629-Centricity Universal Viewer Zero Footprint / ZFP-01300332; 2) 2089507-093, Installed Product ID Numbers: 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, Installed Product ID Numbers: JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, Installed Product ID Numbers: 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, Installed Product ID Numbers: 6) 2089507-112, Installed Product ID Numbers: 7) 2089507-114, Installed Product ID Numbers: 8) 2089507-115, Installed Product ID Numbers: 9) 2089507-116, Installed Product ID Numbers: 10) 2089507-117, Installed Product ID Numbers: 11) 2089507-118, Installed Product ID Numbers: 12) 2089507-119, Installed Product ID Numbers:
Quantity recalled120 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0318-2025
FDA event ID
95512
Recalling firm
GE Healthcare, Chicago, IL
Classification
Class II
Status
Ongoing
Recall initiated
3 Oct 2024
FDA classified
7 Nov 2024
Reported
13 Nov 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

3 Oct 2024Recall initiated by company
7 Nov 2024Classified by FDA+35 days
13 Nov 2024Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.

Product as listed by the FDA

GE Healthcare Centricity Universal Viewer Zero Footprint, Model Numbers: 1) 2089507-083, GDZH-JINWAN-Centricity 2) 2089507-093, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, 105468-1-Centricity Universal Viewer Zero Footprint / ZFP-02254840, 1495-3-Centricity Universal Viewer Zero Footprint / ZFP-00096090, JP-340111-Centricity Universal Viewer Zero Footprint / ZFP-01275379, JP-275353-Centricity Universal Viewer Zero Footprint / ZFP-02042636; 6) 2089507-112, 3922-25-Centricity Universal Viewer Zero Footprint / ZFP-01290685, 3922-17-Centricity Universal Viewer Zero Footprint / ZFP-00098229, 1503586-Centricity Universal Viewer Zero Footprint / ZFP-21016603, RU0994-Centricity Universal Viewer Zero Footprint / ZFP-02331255; 7) 2089507-114, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-02360229, 1461-1-Centricity Universal Viewer Zero Footprint / ZFP-01803680, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225897, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225898, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01225904, 3728-2-Centricity Universal Viewer Zero Footprint / ZFP-01371500, M51725-Centricity Universal Viewer Zero Footprint / ZFP-01300873, M1781854-Centricity Universal Viewer Zero Footprint / ZFP-01553772, 609334-Centricity Universal Viewer Zero Footprint / ZFP-01282257, UJJF01-Centricity Universal Viewer Zero Footprint / ZFP-22186138, CMC-Centricity Universal Viewer Zero Footprint / ZFP-01820241, JP-100014-Centricity Universal Viewer Zero Footprint / ZFP-01890709, KW1102-Centricity Universal Viewer Zero Footprint / ZFP-02191178, 59219_ERPerror-Centricity Universal Viewer Zero Footprint / ZFP-02327073, QA1114-Centricity Universal Viewer Zero Footprint / ZFP-02306397, SAU1433-Centricity Universal Viewer Zero Footprint / ZFP-01457278, TTSH-CVIS-Centricity Universal Viewer Zero Footprint / ZFP-02041523, NE4651-Centricity Universal Viewer Zero Footprint / ZFP-00038599, NE6703-Centricity Universal Viewer Zero Footprint / ZFP-00038552, NE6731-Centricity Universal Viewer Zero Footprint / ZFP-00038461, NE3006-Centricity Universal Viewer Zero Footprint / ZFP-00038362, NE2196-Centricity Universal Viewer Zero Footprint / ZFP-00038588, NPV738-Centricity Universal Viewer Zero Footprint / ZFP-00038637, NE2031-Centricity Universal Viewer Zero Footprint / ZFP-00038578, NE7405-Centricity Universal Viewer Zero Footprint / ZFP-00038372, NE2162-Centricity Universal Viewer Zero Footprint / ZFP-00038563, NE4878-

Where it was distributed

Worldwide distribution - United States Nationwide and the countries of Algeria, Argentina, Brazil, China, Dominican Republic, France, Germany, India, Indonesia, Ireland, Israel, Japan, Kuwait, Malaysia, Mexico, Panama, Qatar, Russia, Saudi Arabia, Singapore, Spain, United Kingdom.

Nationwide

Questions about this recall

Why was it recalled?

GE HealthCare has become aware of an issue in Centricity Universal Viewer Zero Footprint Client (ZFP) where there is a potential security vulnerability which could allow a malicious actor to access the system and potentially manipulate patient data.

Which lots are affected?

UDI/DI 00840682102988 for all models: 1) 2089507-083, Installed Product ID Numbers: GDZH-JINWAN-Centricity Universal Viewer Zero Footprint / ZFP-02171477, 103378-Centricity Universal Viewer Zero Footprint / ZFP-20039058, 1664823-Centricity Universal Viewer Zero Footprint / ZFP-20080654, 493356-Centricity Universal Viewer Zero Footprint / ZFP-20085691, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291850, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02291877, Medireport-Centricity Universal Viewer Zero Footprint / ZFP-02292580, 81077-Centricity Universal Viewer Zero Footprint / ZFP-20680055, IL1469-Centricity Universal Viewer Zero Footprint / ZFP-02156638, JP-110145-Centricity Universal Viewer Zero Footprint / ZFP-01275393, 1333629-Centricity Universal Viewer Zero Footprint / ZFP-01300332; 2) 2089507-093, Installed Product ID Numbers: 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02268286, 10510-3-Centricity Universal Viewer Zero Footprint / ZFP-02030900, 1495-1-Centricity Universal Viewer Zero Footprint / ZFP-02347996, 423834-Centricity Universal Viewer Zero Footprint / ZFP-02286658, 1407636-Centricity Universal Viewer Zero Footprint / ZFP-02304121, 651240-Centricity Universal Viewer Zero Footprint / ZFP-02173181, 268034-Centricity Universal Viewer Zero Footprint / ZFP-02402520; 3) 2089507-098, Installed Product ID Numbers: JP-310027-Centricity Universal Viewer Zero Footprint / ZFP-02033125; 4) 2089507-109, Installed Product ID Numbers: 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095312, 3909-2-Centricity Universal Viewer Zero Footprint / ZFP-20095313, 327750-Centricity Universal Viewer Zero Footprint / ZFP-02184508; 5) 2089507-111, Installed Product ID Numbers: 6) 2089507-112, Installed Product ID Numbers: 7) 2089507-114, Installed Product ID Numbers: 8) 2089507-115, Installed Product ID Numbers: 9) 2089507-116, Installed Product ID Numbers: 10) 2089507-117, Installed Product ID Numbers: 11) 2089507-118, Installed Product ID Numbers: 12) 2089507-119, Installed Product ID Numbers:

Where was it sold?

Worldwide distribution - United States Nationwide and the countries of Algeria, Argentina, Brazil, China, Dominican Republic, France, Germany, India, Indonesia, Ireland, Israel, Japan, Kuwait, Malaysia, Mexico, Panama, Qatar, Russia, Saudi Arabia, Singapore, Spain, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.