CLASS II ONGOING Device recall · Reported 6 Nov 2024 · FDA Enforcement Report

ClariTEE Probe- Miniature transesophageal echo ultrasound recalled

ImaCor is recalling ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor, distributed in Florida, Georgia, New Jersey, Tennessee. The recall was initiated on 16 Sep 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI - Primary - 00861589000108 Lot numbers: 20250509 and 20250823
Quantity recalled73 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0231-2025
FDA event ID
95459
Recalling firm
ImaCor, Jericho, NY
Classification
Class II
Status
Ongoing
Recall initiated
16 Sep 2024
FDA classified
25 Oct 2024
Reported
6 Nov 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

16 Sep 2024Recall initiated by company
25 Oct 2024Classified by FDA+39 days
6 Nov 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

ClariTEE Probe- Miniature transesophageal echo ultrasound probe for use with ImaCor ultrasound diagnostic platforms. Model/Catalog Number: CLT-010, CLT-010-1

Where it was distributed

FL ,GA, NJ, TN

FloridaGeorgiaNew JerseyTennessee

Questions about this recall

Why was it recalled?

Epoxy used to seal the edge of the transducer into its paddle exceeded specification. This variability could attenuate the ultrasound signal with a potential impact on the quality of images obtained from the probes.

Which lots are affected?

UDI - Primary - 00861589000108 Lot numbers: 20250509 and 20250823

Where was it sold?

FL ,GA, NJ, TN

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.