CLASS I ONGOING Device recall · Reported 28 Oct 2026 · FDA Enforcement Report

Philips Lumify Diagnostic Ultrasound System recalled

Philips Ultrasound is recalling Philips Lumify Diagnostic Ultrasound System, distributed nationwide in the US. The recall was initiated on 24 Sep 2026 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesUDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.X
Quantity recalled13,206

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0152-2027
FDA event ID
99878
Recalling firm
Philips Ultrasound, Bothell, WA
Classification
Class I
Status
Ongoing
Recall initiated
24 Sep 2026
FDA classified
16 Oct 2026
Reported
28 Oct 2026
Type
Voluntary: Firm initiated
Reason category
Other Reasons

Timeline

24 Sep 2026Recall initiated by company
16 Oct 2026Classified by FDA+22 days
28 Oct 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.

Product as listed by the FDA

Philips Lumify Diagnostic Ultrasound System

Where it was distributed

Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.

Which lots are affected?

UDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.X

Where was it sold?

Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 3 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Philips Ultrasound

All 110

Other Imaging Equipment recalls

All 2,349