CLASS II Device recall · Reported 9 Jul 2014 · FDA Enforcement Report
CNS-6200 Series Central Nurse Station and accessories recalled
Nihon Kohden America is recalling CNS-6200 Series Central Nurse Station and accessories, distributed in Oklahoma. The recall was initiated on 3 Jun 2014 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1979-2014
- FDA event ID
- 68459
- Recalling firm
- Nihon Kohden America, Irvine, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 3 Jun 2014
- FDA classified
- 2 Jul 2014
- Reported
- 9 Jul 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Patient Monitors
Timeline
| 3 Jun 2014 | Recall initiated by company | |
| 2 Jul 2014 | Classified by FDA | +29 days |
| 9 Jul 2014 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.
Where it was distributed
Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.
Questions about this recall
Why was it recalled?
Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.
Which lots are affected?
Model Number: CNS-6201 Software Version 02-26 only. Serial Numbers: 00485, 00488, 00489, 00490, 00491, 00361, 00429, 00430, 00432, 00436, 00443, 00447, 00478, 00486, 00546, 00564, 00573, 00574, 00575, 00576, 00592, 00626, 00638, 00643, 00651, 00674, 00675, 00677, 00678, 00679, 00687.
Where was it sold?
Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Nihon Kohden America
All 10| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Adult Cap-ONE Biteblock recalled by Nihon Kohden America | Nihon Kohden America | Other | Nationwide |
| Device | CLASS II | Nihon Kohden Adult/Pediatric Forehead Disposable SpO2 Sensor, 0.9m recalled | Nihon Kohden America | Device Malfunction | 29 states |
| Device | CLASS II | Nihon Kohden Adult Ear Clip SpO2 Sensor, 1.5 meter recalled | Nihon Kohden America | Device Malfunction | 29 states |
| Device | CLASS II | BSM-3000 Series Bedside Monitor recalled by Nihon Kohden America | Nihon Kohden America | Device Malfunction | Nationwide |
| Device | CLASS II | WMTS Telemetry Receiver, Multiple Patient Receiver recalled | Nihon Kohden America | Other | 46 states |
Other Patient Monitors recalls
All 1,013| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Bravo CF capsule delivery device, Product Number FGS recalled | Given Imaging | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | Lux-Dx, Lux-Dx II, Lux-Dx II+ Insertable Cardiac Monitor recalled | Boston Scientific | Device Malfunction | Nationwide |
| Device | CLASS II | IQecg Digital ECG recalled by Midmark | Midmark | Labeling Errors | Nationwide |