CLASS II Device recall · Reported 9 Jul 2014 · FDA Enforcement Report

CNS-6200 Series Central Nurse Station and accessories recalled

Nihon Kohden America is recalling CNS-6200 Series Central Nurse Station and accessories, distributed in Oklahoma. The recall was initiated on 3 Jun 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number: CNS-6201 Software Version 02-26 only. Serial Numbers: 00485, 00488, 00489, 00490, 00491, 00361, 00429, 00430, 00432, 00436, 00443, 00447, 00478, 00486, 00546, 00564, 00573, 00574, 00575, 00576, 00592, 00626, 00638, 00643, 00651, 00674, 00675, 00677, 00678, 00679, 00687.
Quantity recalled430

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1979-2014
FDA event ID
68459
Recalling firm
Nihon Kohden America, Irvine, CA
Classification
Class II
Status
Terminated
Recall initiated
3 Jun 2014
FDA classified
2 Jul 2014
Reported
9 Jul 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

3 Jun 2014Recall initiated by company
2 Jul 2014Classified by FDA+29 days
9 Jul 2014Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

CNS-6200 Series Central Nurse Station and accessories. Model Number: CNS-6201 For cardiac and vital signs monitoring for multiple patients.

Where it was distributed

Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.

Oklahoma

Questions about this recall

Why was it recalled?

Software Version 02-26, when used with the Central Nursing Station (CNS) 6201, (PU-621 RA) may unexpectedly and without warning reboot, resulting in a period of approximately 3 minutes during which time the system does not display patient data and does not emit alarms based on that data should they be appropriate.

Which lots are affected?

Model Number: CNS-6201 Software Version 02-26 only. Serial Numbers: 00485, 00488, 00489, 00490, 00491, 00361, 00429, 00430, 00432, 00436, 00443, 00447, 00478, 00486, 00546, 00564, 00573, 00574, 00575, 00576, 00592, 00626, 00638, 00643, 00651, 00674, 00675, 00677, 00678, 00679, 00687.

Where was it sold?

Worldwide Distribution - US in the state of Oklahoma and in the countries of Bolivia, Chile, Germany, Japan, Korea, Taiwan, Yemen, and South Africa.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Nihon Kohden America

All 10

Other Patient Monitors recalls

All 1,013