CLASS II ONGOING Device recall · Reported 17 Aug 2022 · FDA Enforcement Report

cobas e801 Immunoassay Analyzer recalled by Roche Diagnostics Operations

Roche Diagnostics Operations is recalling cobas e801 Immunoassay Analyzer, distributed nationwide in the US. The recall was initiated on 14 Jul 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 07613336158456 and 04015630946198; Catalog No. 08454345001 and 07682913001. All serial numbers.
Quantity recalled965 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1515-2022
FDA event ID
90657
Recalling firm
Roche Diagnostics Operations, Indianapolis, IN
Classification
Class II
Status
Ongoing
Recall initiated
14 Jul 2022
FDA classified
11 Aug 2022
Reported
17 Aug 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

14 Jul 2022Recall initiated by company
11 Aug 2022Classified by FDA+28 days
17 Aug 2022Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo IgG.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

cobas e801 Immunoassay Analyzer

Where it was distributed

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, DC & PR.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland

Questions about this recall

Is cobas e801 Immunoassay Analyzer recalled?

Yes. Roche Diagnostics Operations recalled cobas e801 Immunoassay Analyzer on 14 Jul 2022. The recall is Class II and its status is Ongoing. Recall number Z-1515-2022.

Why was it recalled?

Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo IgG.

Which lots are affected?

UDI-DI: 07613336158456 and 04015630946198; Catalog No. 08454345001 and 07682913001. All serial numbers.

Where was it sold?

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, DC & PR.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Roche Diagnostics Operations

All 127

Other Test Kits & Diagnostics recalls

All 2,656