CLASS II Device recall · Reported 12 Dec 2012 · FDA Enforcement Report
Cobas 8000 Modular Analyzer Series recalled
Roche Diagnostics Operations is recalling Cobas 8000 Modular Analyzer Series, distributed nationwide in the US. The recall was initiated on 1 Nov 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0473-2013
- FDA event ID
- 63577
- Recalling firm
- Roche Diagnostics Operations, Indianapolis, IN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 1 Nov 2012
- FDA classified
- 4 Dec 2012
- Reported
- 12 Dec 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Lead & Heavy Metals
- Device type
- Test Kits & Diagnostics
Timeline
| 1 Nov 2012 | Recall initiated by company | |
| 4 Dec 2012 | Classified by FDA | +33 days |
| 12 Dec 2012 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C
Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.
Product as listed by the FDA
Cobas 8000 Modular Analyzer Series, Model/Catalog/Part Number: 05641489001. The Cobas 8000 is a fully automated system for clinical chemistry analysis.
Where it was distributed
Nationwide Distribution, including the states of AL, AZ, CA, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, NJ, NY, OH, VA, WA, and WI.
Nationwide AlabamaArizonaCaliforniaFloridaGeorgiaIllinoisIndianaKansasKentuckyMassachusettsMarylandMichiganMinnesotaNorth CarolinaNew JerseyNew YorkOhioVirginiaWashingtonWisconsin
Questions about this recall
Is Cobas 8000 Modular Analyzer Series recalled?
Yes. Roche Diagnostics Operations recalled Cobas 8000 Modular Analyzer Series on 1 Nov 2012. The recall is Class II and its status is Terminated. Recall number Z-0473-2013.
Why was it recalled?
On cobas c 701 and cobas c 702 modules of the Cobas 8000 series analyzer, a mismatch between the reagent probe positioning and the aspirated reagent volumes under certain conditions can lead to insufficient reagent pipetting. This may affect patient sample and controls recovery for the following tests: Alkaline Phosphatase (ALP2), Alanine Aminotransferase (ALT), Bicarbonate (CO2-L), Creatinine (C
Which lots are affected?
Catalog numbers affected: Alkaline Phosphatase acc. to IFCC Gen.2 05166888190; Alanine Aminotransferase acc. to IFCC without pyridoxal phosphate (P5P) activation 05850797190; Bicarbonate Liquid 05446376190; Creatinine Jaff¿ Gen.2 05168597190; C-Reactive Protein Gen.3 05172373190; and Urea/BUN 05171873190. Serial Number(s): 0917-10, 1020-04, 1020-05, 1021-03, 1021-04, 1021-07, 1021-08, 1023-06, 1023-09, 1023-10, 1024-03, 1024-04, 1024-05, 1024-09, 1025-02, 1025-05, 1025-07, 1025-09, 1026-01, 1026-02, 1027-09, 1028-01, 1028-07, 1030-09, 1031-05, 1031-09, 1032-09, 1032-10, 1033-04, 1033-05, 1107-06, 1109-02, 1109-03, 1111-07, 1111-08, 1112-02, 1112-04, 1115-06, 1115-07, 1115-08, 1115-10, 1117-04, 1117-07, 1118-02, 1126-02, 1126-03, 1126-04, 1127-01, 1127-02, 1127-06, 1127-07, 1138-03, 1138-04, 1139-04, 1139-05, 1139-06, 1140-02, 1140-03, 1140-04, 1140-05, 1140-10, 1141-08, 1141-09, 1142-02, 1142-03, 1142-04, 1143-01, 1143-02, 1143-03, 1143-04, 1143-05, 1143-06, 1144-05, 1146-04, 1146-05, 1146-06, 1146-07, and 1146-08.
Where was it sold?
Nationwide Distribution, including the states of AL, AZ, CA, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, NJ, NY, OH, VA, WA, and WI.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Roche Diagnostics Operations
All 127| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | cobas pro integrated solutions recalled by Roche Diagnostics Operations | Roche Diagnostics Operations | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | The Elecsys Anti-TSHR immunoassay is a three-step recalled | Roche Diagnostics Operations | Particulate Matter | Nationwide |
| Device | CLASS II | cobas HCYS, Homocysteine Enzymatic Assay, Material Number recalled | Roche Diagnostics Operations | Potency & Assay | 8 states |
| Device | CLASS II | Creatine Kinase (CK) used on cobas c 311/501/502 recalled | Roche Diagnostics Operations | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Tina-quant D-Dimer Test System recalled by Roche Diagnostics Operations | Roche Diagnostics Operations | Potency & Assay | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein S recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Lead & Heavy Metals | Nationwide |