CLASS II Device recall · Reported 2 Sep 2020 · FDA Enforcement Report

Cobra x5 2-Ch EMG ET Tube recalled by Neurovision Medical Products

Neurovision Medical Products is recalling Cobra x5 2-Ch EMG ET Tube, distributed nationwide in the US. The recall was initiated on 21 Jul 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesKit Lots: 021919D 022019B 040119C 040319C 040419A 071019A 010820C 020520C 021020I 021220A 061220A 061720C 120419E 120919B 042319A 072419B 091019A 091619A 091719C 092019A 092419A 050219B 051419B 052819A 060519B 070519A 022019B 022619C 061319B 061419B 072619B 061820C 080919A 081419C 082219A 082619C 052120A 110719A 111919A 121319A 022020B 022420A 031720C 062420C 041020A
Quantity recalled3,425

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2907-2020
FDA event ID
86102
Recalling firm
Neurovision Medical Products, Ventura, CA
Classification
Class II
Status
Terminated
Recall initiated
21 Jul 2020
FDA classified
27 Aug 2020
Reported
2 Sep 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

21 Jul 2020Recall initiated by company
27 Aug 2020Classified by FDA+37 days
2 Sep 2020Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

Product as listed by the FDA

Cobra x5 2-Ch EMG ET Tube, 7.0mm, REF: LTE700DCM-5, Rx Only, Sterile EO UDI: B006LTE700DCM52 - Product Usage: For continuous EMG monitoring and status assessment of the nerves supplying the laryngeal musculature as well as providing an open airway for patient ventilation during surgical or parasurgical care using any compatible nerve monitoring system.

Where it was distributed

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIllinoisIndianaKansasMassachusettsMarylandMichiganMinnesotaMissouriMississippiNebraska

Questions about this recall

Why was it recalled?

The electrode insulating material on the Cobra EMG ET tube used for continuous EMG monitoring may soften over time so that it may be more susceptible to damage by scratching or bending, which could stop or inhibit the EMG signal transmission of the electrodes to the monitor and render the device not usable.

Which lots are affected?

Kit Lots: 021919D 022019B 040119C 040319C 040419A 071019A 010820C 020520C 021020I 021220A 061220A 061720C 120419E 120919B 042319A 072419B 091019A 091619A 091719C 092019A 092419A 050219B 051419B 052819A 060519B 070519A 022019B 022619C 061319B 061419B 072619B 061820C 080919A 081419C 082219A 082619C 052120A 110719A 111919A 121319A 022020B 022420A 031720C 062420C 041020A

Where was it sold?

Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, MA, MD, MI, MN, MO, MS, NE, NY, OH, OR, PA, SC, SD, TN, TX, WA, WI and the countries of Canada, Dominican Republic, Hong Kong.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

8 other products · FDA event 86102

This product is part of a recall event; the main page for the event is Nerveana x5 RLN Monitoring Kit with Cobra 1-Ch EMG ET Tube recalled.

More recalls from Neurovision Medical Products

All 11

Other Surgical Instruments recalls

All 3,809