CLASS II ONGOING Device recall · Reported 7 Jul 2021 · FDA Enforcement Report

Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI recalled

Philips North America is recalling Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly, distributed nationwide in the US. The recall was initiated on 1 May 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel/ Serial Number/UDI: 728120 2516 N/A 728120 3179 N/A 728120 3316 N/A 728201 9326 N/A 728120 3304 N/A 728120 3271 N/A 728120 2537 N/A 728120 3324 N/A 728120 H3321 N/A 728120 3058 N/A 728125 2529 N/A
Quantity recalled1,301 Monitor Bracket Assemblies total

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1979-2021
FDA event ID
87974
Recalling firm
Philips North America, Cambridge, MA
Classification
Class II
Status
Ongoing
Recall initiated
1 May 2021
FDA classified
30 Jun 2021
Reported
7 Jul 2021
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals

Timeline

1 May 2021Recall initiated by company
30 Jun 2021Classified by FDA+60 days
7 Jul 2021Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems: Model - MX8000 Models: 728120/728125/728201

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.

Nationwide ArkansasColoradoDelawareIdahoIllinoisIndianaLouisianaMinnesotaPennsylvaniaTennessee

Questions about this recall

Is Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly recalled?

Yes. Philips North America recalled Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly on 1 May 2021. The recall is Class II and its status is Ongoing. Recall number Z-1979-2021.

Why was it recalled?

Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

Which lots are affected?

Model/ Serial Number/UDI: 728120 2516 N/A 728120 3179 N/A 728120 3316 N/A 728201 9326 N/A 728120 3304 N/A 728120 3271 N/A 728120 2537 N/A 728120 3324 N/A 728120 H3321 N/A 728120 3058 N/A 728125 2529 N/A

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE, AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Philips North America

All 531

Other Imaging Equipment recalls

All 2,336