CLASS II Device recall · Reported 5 Sep 2018 · FDA Enforcement Report

Cortisol Saliva Luminescence Immunoassay The IBL recalled by Tecan US

Tecan US is recalling Cortisol Saliva Luminescence Immunoassay The IBL International Cortisol Saliva, distributed in 9 states. The recall was initiated on 12 Jan 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesIn USA RE62119 Lot # LCL129 exp. date: 2017-11-30, RE62119 Lot # LCL133 exp. date: 2019-03-31, and RE62111 Lot # LCL130 exp. date: 2018-03-31. OUS Lot 131, Lot 132
Quantity recalled870 (USA) 180 (OUS)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2923-2018
FDA event ID
80554
Recalling firm
Tecan US, Morrisville, NC
Classification
Class II
Status
Terminated
Recall initiated
12 Jan 2018
FDA classified
24 Aug 2018
Reported
5 Sep 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

12 Jan 2018Recall initiated by company
24 Aug 2018Classified by FDA+224 days
5 Sep 2018Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Cortisol Saliva Luminescence Immunoassay The IBL International Cortisol Saliva Luminescence immunoassay is intended for the in-vitro diagnostic quantitative determination of Cortisol in human saliva and for use as an aid in the diagnosis and treatment of adrenal disorders. The device is not intended for point-of-care settings.

Where it was distributed

Worldwide distribution; US distribution to states of: GA, IA, LA, MA, MN, OR, PA, SC, and UT; and countries of: Germany, Finland, Switzerland, France, Austria, Denmark, Poland, Norway, Belgium, Netherlands, Italy, Canada, South Africa, Russia, and Japan.

GeorgiaIowaLouisianaMassachusettsMinnesotaOregonPennsylvaniaSouth CarolinaUtah

Questions about this recall

Why was it recalled?

Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard.

Which lots are affected?

In USA RE62119 Lot # LCL129 exp. date: 2017-11-30, RE62119 Lot # LCL133 exp. date: 2019-03-31, and RE62111 Lot # LCL130 exp. date: 2018-03-31. OUS Lot 131, Lot 132

Where was it sold?

Worldwide distribution; US distribution to states of: GA, IA, LA, MA, MN, OR, PA, SC, and UT; and countries of: Germany, Finland, Switzerland, France, Austria, Denmark, Poland, Norway, Belgium, Netherlands, Italy, Canada, South Africa, Russia, and Japan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.