CLASS II Device recall · Reported 17 Apr 2013 · FDA Enforcement Report

Dade(R) Actin(R) FSL Activated PTT Reagent recalled

Siemens Healthcare Diagnostics is recalling Dade(R) Actin(R) FSL Activated PTT Reagent, distributed nationwide in the US. The recall was initiated on 11 Mar 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog numbers B4219-1 and B4219-2: lot numbers 547311, exp 3/22/2013; 547312, exp 3/22/2013 and 547316A, exp 4/28/2013.
Quantity recalled13,083

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1073-2013
FDA event ID
63804
Recalling firm
Siemens Healthcare Diagnostics, Newark, DE
Classification
Class II
Status
Terminated
Recall initiated
11 Mar 2013
FDA classified
9 Apr 2013
Reported
17 Apr 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

11 Mar 2013Recall initiated by company
9 Apr 2013Classified by FDA+29 days
17 Apr 2013Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high.

Product as listed by the FDA

Dade(R) Actin(R) FSL Activated PTT Reagent. For use in the determination of the activated partial thromboplastin time (APTT).

Where it was distributed

Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV and Puerto Rico and the country of Canada.

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaine

Questions about this recall

Why was it recalled?

Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high.

Which lots are affected?

Catalog numbers B4219-1 and B4219-2: lot numbers 547311, exp 3/22/2013; 547312, exp 3/22/2013 and 547316A, exp 4/28/2013.

Where was it sold?

Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV and Puerto Rico and the country of Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.