CLASS II ONGOING Device recall · Reported 11 Sep 2024 · FDA Enforcement Report

Desflurane E_Vap.(Saf-T-Fill) used recalled by Mindray DS USA

Mindray DS USA is recalling Desflurane E_Vap.(Saf-T-Fill) used with the A9 Anesthesia System-Used to provide the, distributed in 8 states. The recall was initiated on 22 Jul 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI:06936415975760 Serial Number: AH6C25000263 AH6C25000255 AH6C27000355 AH6C27000346 AH6C27000371 AH6C25000256 AH6C25000296 AH6C25000290 AH6C27000361 AH6C25000254 AH6C25000254 AH6C25000271 AH6C25000298 AH6C25000299 AH6C25000300 AH6C25000301 AH6C25000302 AH6C27000339 AH6C27000343 AH6C27000344 AH6C27000347 AH6C27000348 AH6C27000350 AH6C27000352 AH6C27000354 AH6C27000358 AH6C27000359 AH6C27000363 AH6C27000364 AH6C27000366 AH6C27000410 AH6C27000411 AH6C28000444 AH6C28000447 AH6C28000448 AH6C28000449 AH6C28000450 AH6C28000454
Quantity recalled96 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2952-2024
FDA event ID
95099
Recalling firm
Mindray DS USA, Mahwah, NJ
Classification
Class II
Status
Ongoing
Recall initiated
22 Jul 2024
FDA classified
4 Sep 2024
Reported
11 Sep 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Packaging Defects

Timeline

22 Jul 2024Recall initiated by company
4 Sep 2024Classified by FDA+44 days
11 Sep 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Desflurane E_Vap.(Saf-T-Fill) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: Part Number115-066759-00

Where it was distributed

US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

CaliforniaGeorgiaKentuckyMississippiNew JerseyPuerto RicoTexasWest Virginia

Questions about this recall

Why was it recalled?

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Which lots are affected?

UDI-DI:06936415975760 Serial Number: AH6C25000263 AH6C25000255 AH6C27000355 AH6C27000346 AH6C27000371 AH6C25000256 AH6C25000296 AH6C25000290 AH6C27000361 AH6C25000254 AH6C25000254 AH6C25000271 AH6C25000298 AH6C25000299 AH6C25000300 AH6C25000301 AH6C25000302 AH6C27000339 AH6C27000343 AH6C27000344 AH6C27000347 AH6C27000348 AH6C27000350 AH6C27000352 AH6C27000354 AH6C27000358 AH6C27000359 AH6C27000363 AH6C27000364 AH6C27000366 AH6C27000410 AH6C27000411 AH6C28000444 AH6C28000447 AH6C28000448 AH6C28000449 AH6C28000450 AH6C28000454

Where was it sold?

US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

3 other products · FDA event 95099

This product is part of a recall event; the main page for the event is Sevoflurane E_Vap.(Safety Filling adapter) used recalled.

More recalls from Mindray DS USA

All 47

Other Other Medical Devices recalls

All 13,950