CLASS II Device recall · Reported 3 Oct 2012 · FDA Enforcement Report

Dexis Imaging Suite recalled by Imaging Sciences International

Imaging Sciences International is recalling Dexis Imaging Suite, distributed in 42 states. The recall was initiated on 30 Jul 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog number 1.008.7932, version numbers 10.0.0 through 10.0.3
Quantity recalled366

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2463-2012
FDA event ID
62830
Recalling firm
Imaging Sciences International, Hatfield, PA
Classification
Class II
Status
Terminated
Recall initiated
30 Jul 2012
FDA classified
26 Sep 2012
Reported
3 Oct 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

30 Jul 2012Recall initiated by company
26 Sep 2012Classified by FDA+58 days
3 Oct 2012Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A condition can occur during renumbering of the patient database using the DEXCopy feature. When renumbering is being performed using the DEXCopy feature and multiple patients with an identical name are present, the software will assign all subsequent records to the first patient record, resulting in mixed patient records. In order for this problem to occur several conditions need to be met: 1)

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

DEXIS Imaging Suite - catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3 Radiological Image Processing System

Where it was distributed

Worldwide Distribution - USA including Washington, DC, AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and the country of Canada.

AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaIowaIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichiganMinnesota

Questions about this recall

Is Dexis Imaging Suite recalled?

Yes. Imaging Sciences International recalled Dexis Imaging Suite on 30 Jul 2012. The recall is Class II and its status is Terminated. Recall number Z-2463-2012.

Why was it recalled?

A condition can occur during renumbering of the patient database using the DEXCopy feature. When renumbering is being performed using the DEXCopy feature and multiple patients with an identical name are present, the software will assign all subsequent records to the first patient record, resulting in mixed patient records. In order for this problem to occur several conditions need to be met: 1)

Which lots are affected?

Catalog number 1.008.7932, version numbers 10.0.0 through 10.0.3

Where was it sold?

Worldwide Distribution - USA including Washington, DC, AL, AZ, AR, CA, CO, CT, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and the country of Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.