CLASS II Device recall · Reported 27 May 2020 · FDA Enforcement Report

Dfine Vertebral Hydraulic Assemblies: Hydraulic Assembly recalled

Merit Medical Systems is recalling Dfine Vertebral Hydraulic Assemblies: Hydraulic Assembly, StabiliT Vertebral Augmentation, distributed nationwide in the US. The recall was initiated on 13 Feb 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot number: AAD-1801-16 and UDI: (01)00884450374399(17)221231(10)AAD-1801-16; Lot number: AAD-1801-17 and UDI: (01)00884450374399(17)221231(10)AAD-1801-17; Lot number: AAD-1803-01 and UDI: (01)00884450374399(17)230331(10)AAD-1803-01; Lot number: AAD-1804-05 and UDI: (01)00884450374399(17)230430(10)AAD-1804-05; Lot number: AAD-1901-07 and UDI: (01)00884450374399(17)231231(10)AAD-1901-07; Lot number: AAD-1904-04 and UDI: (01)00884450374399(17)240430(10)AAD-1904-04.
Quantity recalled228

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2067-2020
FDA event ID
85485
Recalling firm
Merit Medical Systems, South Jordan, UT
Classification
Class II
Status
Terminated
Recall initiated
13 Feb 2020
FDA classified
21 May 2020
Reported
27 May 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Spinal Implants

Timeline

13 Feb 2020Recall initiated by company
21 May 2020Classified by FDA+98 days
27 May 2020Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed on the labels is five (5) years.The affected product will begin expiring in March 2020. Using product beyond the validated shelf-life may result in (1) product not functioning as intended and/or (2) a delay in procedure or the need for the procedure to be rescheduled.

Product as listed by the FDA

DFINE Vertebral Hydraulic Assemblies: Hydraulic Assembly, StabiliT Vertebral Augmentation System, Reference/Catalog number 1402 - Product Usage: is intended for percutaneous delivery of StabiliT ER2 Bone Cement in kyphoplasty procedures in the treatment of pathological fractures of the vertebrae. Painful vertebral compression fractures may result from osteoporosis, benign lesions (hemangioma), and malignant lesions (metastatic cancers, myeloma).

Where it was distributed

Worldwide distribution: US Nationwide distribution including in the states of (TX, UT, PA, KY, TX, AR, CA, and PA) which included one (1) government/military consignee in KY; and the countries of (Germany, Switzerland, and New Zealand).

Nationwide ArkansasCaliforniaKentuckyPennsylvaniaTexasUtah

Questions about this recall

Why was it recalled?

Vertebral hydraulic assemblies has incorrect expiration date printed on the labels.The product's shelf-life is two (2) years, but the expiration date printed on the labels is five (5) years.The affected product will begin expiring in March 2020. Using product beyond the validated shelf-life may result in (1) product not functioning as intended and/or (2) a delay in procedure or the need for the procedure to be rescheduled.

Which lots are affected?

Lot number: AAD-1801-16 and UDI: (01)00884450374399(17)221231(10)AAD-1801-16; Lot number: AAD-1801-17 and UDI: (01)00884450374399(17)221231(10)AAD-1801-17; Lot number: AAD-1803-01 and UDI: (01)00884450374399(17)230331(10)AAD-1803-01; Lot number: AAD-1804-05 and UDI: (01)00884450374399(17)230430(10)AAD-1804-05; Lot number: AAD-1901-07 and UDI: (01)00884450374399(17)231231(10)AAD-1901-07; Lot number: AAD-1904-04 and UDI: (01)00884450374399(17)240430(10)AAD-1904-04.

Where was it sold?

Worldwide distribution: US Nationwide distribution including in the states of (TX, UT, PA, KY, TX, AR, CA, and PA) which included one (1) government/military consignee in KY; and the countries of (Germany, Switzerland, and New Zealand).

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 85485

This product is part of a recall event; the main page for the event is Dfine Vertebral Assemblies: Hydraulic-Master Syringe recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIDfine Vertebral Assemblies: Hydraulic-Master Syringe recalledMerit Medical SystemsOtherNationwide

More recalls from Merit Medical Systems

All 144

Other Spinal Implants recalls

All 1,157