CLASS I Device recall · Reported 1 Apr 2015 · FDA Enforcement Report

Diamondback 360 Coronary Orbital Atherectomy System recalled

Cardiovascular Systems is recalling Diamondback 360 Coronary Orbital Atherectomy System, distributed nationwide in the US. The recall was initiated on 4 Mar 2015 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codes107046, 106992, 106993, 106990.
Quantity recalled30

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1290-2015
FDA event ID
70727
Recalling firm
Cardiovascular Systems, Saint Paul, MN
Classification
Class I
Status
Terminated
Recall initiated
4 Mar 2015
FDA classified
20 Mar 2015
Reported
1 Apr 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

4 Mar 2015Recall initiated by company
20 Mar 2015Classified by FDA+16 days
1 Apr 2015Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The OAD saline sheath outer layer may flake off from the catheter due to friction between the outer material layer and corresponding materials it may come in contact with during clinical use. If the material is large enough, it has the potential to cause a flow limiting embolism.

Product as listed by the FDA

Diamondback 360 Coronary Orbital Atherectomy System, Model Number: DBEC-125. The system is indicated to facilitate stent delivery in patients with coronary artery disease (CAD) who are acceptable candidates for PTCA or stenting due to de novo, severely calcified coronary artery lesions.

Where it was distributed

Nationwide Distribution-including the states of AZ, FL, GA, LA, MA, MS, MO, NJ, NY, OH, and RI.

Nationwide ArizonaFloridaGeorgiaLouisianaMassachusettsMissouriMississippiNew JerseyNew YorkOhioRhode Island

Questions about this recall

Why was it recalled?

The OAD saline sheath outer layer may flake off from the catheter due to friction between the outer material layer and corresponding materials it may come in contact with during clinical use. If the material is large enough, it has the potential to cause a flow limiting embolism.

Which lots are affected?

107046, 106992, 106993, 106990.

Where was it sold?

Nationwide Distribution-including the states of AZ, FL, GA, LA, MA, MS, MO, NJ, NY, OH, and RI.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Cardiovascular Systems

All 15

Other Heart & Vascular Devices recalls

All 6,217