CLASS II ONGOING Device recall · Reported 21 Aug 2024 · FDA Enforcement Report

Discrete photometric chemistry analyzer for clinical use recalled

Beckman Coulter is recalling Discrete photometric chemistry analyzer for clinical use, distributed nationwide in the US. The recall was initiated on 14 Jun 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number: 81600N, C11907; UDI: 150995901; Serial Numbers: 574415, 574440, 574484,574416, 574441, 574485, 574417, 574442, 574486, 574418, 574443, 574488,574419, 574444, 574489,574420, 574445, 574490, 574421, 574446, 574491, 574422, 574447 574494,574423, 574448, 574497,574424, 574454, 574502, 574425, 574455, 574503, 574426, 574456 574504,574427, 574457, 574505,574428, 574459, 574506, 574429, 574469, 574507, 574430, 574470 574509,574432, 574471, 574513, 574433, 574472, 574514, 574434, 574477, 574515, 574435, 574479
Quantity recalled59 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2545-2024
FDA event ID
94902
Recalling firm
Beckman Coulter, Chaska, MN
Classification
Class II
Status
Ongoing
Recall initiated
14 Jun 2024
FDA classified
9 Aug 2024
Reported
21 Aug 2024
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Potency & Assay

Timeline

14 Jun 2024Recall initiated by company
9 Aug 2024Classified by FDA+56 days
21 Aug 2024Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Discrete photometric chemistry analyzer for clinical use.

Where it was distributed

Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.

Nationwide AlabamaCaliforniaGeorgiaIllinoisLouisianaMichiganMontanaNebraskaNew YorkTexas

Questions about this recall

Why was it recalled?

Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.

Which lots are affected?

Catalog Number: 81600N, C11907; UDI: 150995901; Serial Numbers: 574415, 574440, 574484,574416, 574441, 574485, 574417, 574442, 574486, 574418, 574443, 574488,574419, 574444, 574489,574420, 574445, 574490, 574421, 574446, 574491, 574422, 574447 574494,574423, 574448, 574497,574424, 574454, 574502, 574425, 574455, 574503, 574426, 574456 574504,574427, 574457, 574505,574428, 574459, 574506, 574429, 574469, 574507, 574430, 574470 574509,574432, 574471, 574513, 574433, 574472, 574514, 574434, 574477, 574515, 574435, 574479

Where was it sold?

Worldwide - US Nationwide distribution in the states of NE, MT, LA, TX, NY, CA, MI, AL, IL the countries of Austria, Belgium, Georgia, India, Indonesia, Italy, Korea, Malaysia, Myanmar, Nigeria, Switzerland, Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.