CLASS II ONGOING Device recall · Reported 8 Mar 2023 · FDA Enforcement Report
Drx-Compass/Dr-fit X-ray Systems with Firmware recalled
Carestream Health is recalling Drx-Compass/Dr-fit X-ray Systems with Firmware, distributed nationwide in the US. The recall was initiated on 30 Jan 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1206-2023
- FDA event ID
- 91698
- Recalling firm
- Carestream Health, Rochester, NY
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 30 Jan 2023
- FDA classified
- 2 Mar 2023
- Reported
- 8 Mar 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Imaging Equipment
Timeline
| 30 Jan 2023 | Recall initiated by company | |
| 2 Mar 2023 | Classified by FDA | +31 days |
| 8 Mar 2023 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury
Product as listed by the FDA
DRX-Compass/DR-FIT X-ray Systems with Firmware version 2.3.2.0/2.3.2.4. For use in obtaining diagnostic quality radiographic images to aid the physician with diagnosis. CATALOG #(s): 8624181 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8624199 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F 8621617 X-RAY SYSTEM CONSOLE PDU 3P 380-480 C/F 8622359 X-RAY SYSTEM CONSOLE PDU 3P 208-240 C/F Note: PDU is a standalone Power Distribution Unit (PDU), plugged into the hospital outlet, which supplies Alternate Current (AC) to the Overhead Tube Crane (OTC).
Where it was distributed
Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.
Nationwide CaliforniaIdahoIllinoisMassachusettsMinnesotaNorth CarolinaNevadaTexasWisconsin
Questions about this recall
Why was it recalled?
After pressing and releasing the Z-Axis Motorized buttons on the Tube Head Display, the Overhead Tube Crane (OTC) could potentially continue the movement unexpectedly to a certain distance which could potentially result in a injury
Which lots are affected?
UDI: 60889978621613 (Rayco) 60889978624188 (1049) US Serial Numbers: SP0763 50066 50082 50089 50135 50145 50143 50140D 50144 OUS serial numbers: SP0819 SP0687G2 SP0726 SP0746 SP0834 SP0837 S0001 S0170 S0698G2 1085 FS0564 S0578 S0744 SP0903 SP0600 SP1035 SP0930 SP1007 SP1037 SP0772 S0934G2 S0639G2 S1020G2 S0118 S0135 S0134 S0165 S0167 S0168 S0181 S0800G2 S0784G2 S0811G2 S0832G2 S0826G2 S0822G2 S1038G2 S0678G2 S0619G2 S0636G2 S0641G2 S0643G2 S0701G2 S0738G2 S0758G2 S0760G2 S0673G2 S0785G2 S0842G2 S0860G2 S0863G2 S0888G2 S0973G2 S1105G2 S1048G2 S1073G2 S1004G2 S1110G2 S0791G2 S0941G2 SP0900 S0625
Where was it sold?
Worldwide distribution - US Nationwide distribution in the states of California, Idaho, Illinois, Massachusetts, Minnesota, Nevada, North Carolina, Texas Wisconsin and the countries of Argentina, Australia, Brazil, China, Finland, France, INDIA, ITALY, Korea, Malaysia, Poland, Republic of Korea.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Carestream Health
All 86| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DRX Revolution Mobile X-Ray System recalled by Carestream Health | Carestream Health | Device Malfunction | Nationwide |
| Device | CLASS II | The DRX-Revolution Mobile X-Ray System is a diagnostic recalled | Carestream Health | Device Malfunction | Nationwide |
| Device | CLASS II | Kodak DirectView DR3000/3500 recalled by Carestream Health | Carestream Health | Other | IA |
| Device | CLASS II | OnSight 3D Extremity System- X-Ray, Tomography Computed System recalled | Carestream Health | Other | Nationwide |
| Device | CLASS II | DRX-Revolution Mobile X-Ray System recalled by Carestream Health | Carestream Health | Other | Nationwide |
Other Imaging Equipment recalls
All 2,347| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |