DxC 500 AU Clinical Chemistry Analyzer recalled
Beckman Coulter Mishima K.K is recalling DxC 500 AU Clinical Chemistry Analyzer, distributed nationwide in the US. The recall was initiated on 10 Jul 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2691-2024
- FDA event ID
- 95018
- Recalling firm
- Beckman Coulter Mishima K.K, Sunto-Gun
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 10 Jul 2024
- FDA classified
- 21 Aug 2024
- Reported
- 28 Aug 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Test Kits & Diagnostics
Timeline
| 10 Jul 2024 | Recall initiated by company | |
| 21 Aug 2024 | Classified by FDA | +42 days |
| 28 Aug 2024 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Where it was distributed
Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.
Nationwide AlabamaCaliforniaGeorgiaHawaiiIllinoisMichiganMinnesotaMississippiNebraskaNew YorkOregonPennsylvaniaSouth Carolina
Questions about this recall
Why was it recalled?
Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
Which lots are affected?
Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers: 2023070010 2023060005 2023090015 2023070009 2023080011 2023080012 2023100019 2023100018 2023100020 2024010037 2024010036 2023110022 2024010038 2023090016 2024020040 2024020041 2024020039 2024020043 2024020042 2023110023 2024020045 2024020046 2024010024 2024010025
Where was it sold?
Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Beckman Coulter Mishima K.K
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 700 AU recalled by Beckman Coulter Mishima K.K | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Chemistry Analyzer Au5800 recalled by Beckman Coulter Mishima K.K | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Sample probe sucks a sample dispensed into a tube or cup recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Sample probe sucks a sample dispensed into a tube or cup recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
Other Test Kits & Diagnostics recalls
All 2,663| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzers: AU w/out ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules: DxC 500 AU Module w/ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |