CLASS II ONGOING Device recall · Reported 19 Mar 2025 · FDA Enforcement Report

DxI 9000 Access Immunoassay Analyzer recalled by Beckman Coulter

Beckman Coulter is recalling DxI 9000 Access Immunoassay Analyzer, distributed nationwide in the US. The recall was initiated on 13 Nov 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 15099590732103; Serial Numbers: 300116 300117 300123 300124 300126 300128 300129 300130 300131 300132 300133 300134 300135 300136 300137 300138 300139 300141 300142 300143 300144 300145 300147 300148 300149 300150 300151 300152 300153 300154 300155 300156 300157 300158 300161 300162 300163 300164 300166 300167 300168 300169 300170 300171 300172 300173 300174 300176 300177 300178 300179 300181 300182 300183 300186 300188 300189 300190 300191 300192 300193 300194 300195 300197 300198 300199 300200 300201 300202 300203 300204 300205 300206 300207 300208 300209 300210 300211 300212 300213 300214 300215 300216 300217 300218 300219 300220 300221 300222 300223 300224 300225 300226 300227 300228 300229 300230 300231 300232 300233 300234 300235 300236 300237 300238 300239 300240 300241 300242 300243 300244 300245 300246 300247 300248 300250 300251 300252 300253 300254 300255 300256 300257 300258 300259 300260 300261 300263 300264 300266 300267 300268 300269 300270 300271 300272 300273 300274 300275 300276 300277 300279 300280 300281 300282 300283 300284 300285 300286 300287 300288 300289 300290 300291 300293 300294 300296 300297 300298 300299 300300 300301 300302 300303 300304 300305 300306 300307 300308 300309 300312 300313 300314 300315 300316 300317 300318 300319 300320 300321 300322 300324 300325 300326 300328 300329 300330 300331 300332 300333 300334 300336 300337 300338 300339 300340 300341 300343 300346 300349 300350 300351 300352 300353 300354 300355 300356 300358 300360 300361 300362 300366 300367 300368 300370 300371 300372 300373 300375 300376 300377 300378 300379 300380 300381 300382 300385 300388 300389 300391 300392 300393 300395 300396 300397 300398 300399 300400 300402 300403 300404 300405 300406 300407 300408
Quantity recalled245 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1326-2025
FDA event ID
96273
Recalling firm
Beckman Coulter, Chaska, MN
Classification
Class II
Status
Ongoing
Recall initiated
13 Nov 2024
FDA classified
12 Mar 2025
Reported
19 Mar 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

13 Nov 2024Recall initiated by company
12 Mar 2025Classified by FDA+119 days
19 Mar 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137

Where it was distributed

US Nationwide. Global Distribution.

Nationwide

Questions about this recall

Why was it recalled?

The DxI 9000 Access Immunoassay Analyzer wash wheel mixer may not spin freely within the idler pulley housing. When the problem occurs, the analyzer may generate a system event for a Wash Wheel Dispense Arm Spin Mixer Error that is displayed on the navigation bar in the User Interface. The analyzer enters the red system status and will generate a SYS flag. The tests that are in progress may be cancelled and may cause a delay in reporting patient results

Which lots are affected?

UDI-DI: 15099590732103; Serial Numbers: 300116 300117 300123 300124 300126 300128 300129 300130 300131 300132 300133 300134 300135 300136 300137 300138 300139 300141 300142 300143 300144 300145 300147 300148 300149 300150 300151 300152 300153 300154 300155 300156 300157 300158 300161 300162 300163 300164 300166 300167 300168 300169 300170 300171 300172 300173 300174 300176 300177 300178 300179 300181 300182 300183 300186 300188 300189 300190 300191 300192 300193 300194 300195 300197 300198 300199 300200 300201 300202 300203 300204 300205 300206 300207 300208 300209 300210 300211 300212 300213 300214 300215 300216 300217 300218 300219 300220 300221 300222 300223 300224 300225 300226 300227 300228 300229 300230 300231 300232 300233 300234 300235 300236 300237 300238 300239 300240 300241 300242 300243 300244 300245 300246 300247 300248 300250 300251 300252 300253 300254 300255 300256 300257 300258 300259 300260 300261 300263 300264 300266 300267 300268 300269 300270 300271 300272 300273 300274 300275 300276 300277 300279 300280 300281 300282 300283 300284 300285 300286 300287 300288 300289 300290 300291 300293 300294 300296 300297 300298 300299 300300 300301 300302 300303 300304 300305 300306 300307 300308 300309 300312 300313 300314 300315 300316 300317 300318 300319 300320 300321 300322 300324 300325 300326 300328 300329 300330 300331 300332 300333 300334 300336 300337 300338 300339 300340 300341 300343 300346 300349 300350 300351 300352 300353 300354 300355 300356 300358 300360 300361 300362 300366 300367 300368 300370 300371 300372 300373 300375 300376 300377 300378 300379 300380 300381 300382 300385 300388 300389 300391 300392 300393 300395 300396 300397 300398 300399 300400 300402 300403 300404 300405 300406 300407 300408

Where was it sold?

US Nationwide. Global Distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.