CLASS II ONGOING Device recall · Reported 13 Aug 2025 · FDA Enforcement Report

DxI 600 Access Immunoassay Analyzer W/Dual Gantry recalled

Beckman Coulter is recalling DxI 600 Access Immunoassay Analyzer W/Dual Gantry, distributed nationwide in the US. The recall was initiated on 25 Jun 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 15099590369248; Serial Numbers: 902101
Quantity recalled1 system

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2270-2025
FDA event ID
97162
Recalling firm
Beckman Coulter, Chaska, MN
Classification
Class II
Status
Ongoing
Recall initiated
25 Jun 2025
FDA classified
5 Aug 2025
Reported
13 Aug 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

25 Jun 2025Recall initiated by company
5 Aug 2025Classified by FDA+41 days
13 Aug 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461

Where it was distributed

Worldwide distribution - US Nationwide.

Nationwide

Questions about this recall

Why was it recalled?

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

Which lots are affected?

UDI-DI: 15099590369248; Serial Numbers: 902101

Where was it sold?

Worldwide distribution - US Nationwide.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

5 other products · FDA event 97162

This product is part of a recall event; the main page for the event is DxI 800 Access Immunoassay Analyzer W/Dual Gantry recalled.

More recalls from Beckman Coulter

All 334

Other Test Kits & Diagnostics recalls

All 2,658