CLASS II ONGOING Device recall · Reported 26 May 2021 · FDA Enforcement Report

Elecsys Troponin I Immunoassay recalled by Roche Diagnostics Operations

Roche Diagnostics Operations is recalling Elecsys Troponin I Immunoassay, distributed nationwide in the US. The recall was initiated on 7 Apr 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI: 04015630939374
Quantity recalled726 kits

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1596-2021
FDA event ID
87705
Recalling firm
Roche Diagnostics Operations, Indianapolis, IN
Classification
Class II
Status
Ongoing
Recall initiated
7 Apr 2021
FDA classified
14 May 2021
Reported
26 May 2021
Type
Voluntary: Firm initiated
Customers notified by
FAX
Reason category
Potency & Assay

Timeline

7 Apr 2021Recall initiated by company
14 May 2021Classified by FDA+37 days
26 May 2021Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Elecsys Troponin I Immunoassay: Catalog# 05094810160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

Where it was distributed

US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.

Nationwide AlabamaArizonaCaliforniaConnecticutFloridaGeorgiaIdahoIllinoisIndianaKansasKentuckyMarylandMaineMissouriNorth CarolinaNebraskaNew JerseyNew MexicoNew YorkOhio

Questions about this recall

Why was it recalled?

Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.

Which lots are affected?

UDI: 04015630939374

Where was it sold?

US Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 87705

This product is part of a recall event; the main page for the event is Elecsys Troponin I STAT Immunoassay recalled.

ReportedClassRecallCompanyReasonWhere
CLASS IIElecsys Troponin I STAT Immunoassay recalledRoche Diagnostics OperationsPotency & AssayNationwide

More recalls from Roche Diagnostics Operations

All 127

Other Test Kits & Diagnostics recalls

All 2,663