CLASS II Device recall · Reported 3 Sep 2014 · FDA Enforcement Report

Endo & Unipolar Femoral Heads: Endo Femoral Head Dia recalled by Zimmer

Zimmer is recalling Endo & Unipolar Femoral Heads: Endo Femoral Head Dia. The recall was initiated on 22 Jul 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesItem Numbers : 00781804300, 00781805300, 00781804700, 00781805000, 00781804200, 00781805400, 00781804800, 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500 Lot Numbers: 37210958, 37210967, 37211180, 37211183, 37211311, 37211316, 37211320, 37211379, 37211482, 62561484, 37211643, 37211645, 37211646, 37211647, 37211653, 37211655, 37211657, 37211658, 37211661, 37211801, 37211804, 37211805, 37211806, 37211811, 37211812, 37211815, 37211817, 37211820, 37212062, 37212063, 37212065, 37212074, 37212079, 62573574, 62573576, 62576002, 62576008, 62593298, 62593302, 37212608, 37211644
Quantity recalled972,125 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2548-2014
FDA event ID
68818
Recalling firm
Zimmer, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
22 Jul 2014
FDA classified
26 Aug 2014
Reported
3 Sep 2014
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Sterility

Timeline

22 Jul 2014Recall initiated by company
26 Aug 2014Classified by FDA+35 days
3 Sep 2014Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Endo & Unipolar Femoral Heads: ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 41MM DIA, ENDO FEMORAL HEAD 42MM DIA, ENDO FEMORAL HEAD 43MM DIA, ENDO FEMORAL HEAD 44MM DIA, ENDO FEMORAL HEAD 45MM DIA, ENDO FEMORAL HEAD 46MM DIA, ENDO FEMORAL HEAD 47MM DIA, ENDO FEMORAL HEAD 48MM DIA, ENDO FEMORAL HEAD 49MM DIA, ENDO FEMORAL HEAD 50MM DIA, ENDO FEMORAL HEAD 51MM DIA, ENDO FEMORAL HEAD 52MM DIA, ENDO FEMORAL HEAD 53MM DIA, ENDO FEMORAL HEAD 54MM DIA, ENDO FEMORAL HEAD 55MM DIA

Where it was distributed

Worldwide Distribution.

Questions about this recall

Why was it recalled?

Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.

Which lots are affected?

Item Numbers : 00781804300, 00781805300, 00781804700, 00781805000, 00781804200, 00781805400, 00781804800, 00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500 Lot Numbers: 37210958, 37210967, 37211180, 37211183, 37211311, 37211316, 37211320, 37211379, 37211482, 62561484, 37211643, 37211645, 37211646, 37211647, 37211653, 37211655, 37211657, 37211658, 37211661, 37211801, 37211804, 37211805, 37211806, 37211811, 37211812, 37211815, 37211817, 37211820, 37212062, 37212063, 37212065, 37212074, 37212079, 62573574, 62573576, 62576002, 62576008, 62593298, 62593302, 37212608, 37211644

Where was it sold?

Worldwide Distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

6 other products · FDA event 68818

This product is part of a recall event; the main page for the event is Moore & Thompson Hip Stems: Fnstrd Hip 38HDX127ST Fnstrd recalled.

More recalls from Zimmer

All 539

Other Hip & Knee Implants recalls

All 3,097