CLASS II Device recall · Reported 23 Apr 2014 · FDA Enforcement Report

Endosseous Dental Implant recalled by Biomet 3i

Biomet 3i is recalling Endosseous Dental Implant. The recall was initiated on 29 Aug 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number(s): Affected devices catalog names are: BNES310, BNET3210, BNPS6511, BNPS6513, BNPS6585, BNPT6510, BNPT6511, BNPT6513, BNPT6585, BNSS310, BNSS311, BNSS313, BNSS610, BNSS613, BNSS615, BNSS685, BNST3210, BNST3211, BNST3213, BNST3215, BNST3285, BNST611, BNST613, BNST685, BOPS6510, BOPS6511, BOPS6515, BOPS6585, BOPT6510, BOPT6511, BOPT6513, BOPT6585, BOSS310, BOSS311, BOSS313, BOSS315, BOSS611, BOSS613, BOSS685, BOST3210, BOST3211, BOST3213, BOST3215, BOST3285, BOST610, BOST611, BOST613, BOST685, FNT3210, FNT3211, IFNT3210, IFNT3211, IFNT3213, IFNT3215, IFNT3285, IFNT610, IFNT611, IFNT613, IFNT685, IFOSS5611, IFOS5685, IFOS610, IFOS611, IFOS615, IFOSM310, IFOSM311, IFOSM313, IFOSM385, IIOS5610, IIOS5611, INT3210, INT3211, INT3213, INT3215, INT3285, INT610, INT611, INT613, INT685, IOSM311, IOSM313, IOSM315, IOSM385, IOSS610, IOSS613, NIIITP6510, NIIITP6511, NIIITP6513, NIIITP6515, NIIITP6585, NINT3210, NINT3211, NINT3213, NINT3215, NINT610, NINT611, NINT613, NINT685, NIOSM311, NIOSM313, NIOSS610, NT3210, NT3211, OSM310, XIFNT3210, XIFNT3211, XIFNT3213, XIFNT3215, XIFNT3285, XIFN610, XIFNT611, XIFNT613, XIFNT685, XIFOSM310, XIFOSM311, XIFOSM313, XIFOSS610, XIFOSS611, XIIOS6510, XIIOS6511, XIIOS6513, XIITP6510, XIITP6511, XIITP6513. Manufactured dates from November 5, 2012 through August 21, 2013.
Quantity recalled34,744

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1458-2014
FDA event ID
67609
Recalling firm
Biomet 3i, Palm Beach Gardens, FL
Classification
Class II
Status
Terminated
Recall initiated
29 Aug 2013
FDA classified
15 Apr 2014
Reported
23 Apr 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Dental Devices

Timeline

29 Aug 2013Recall initiated by company
15 Apr 2014Classified by FDA+229 days
23 Apr 2014Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to residual machining fluid identified by discoloration (darker in appearance) on the external dental implant surface.

Product as listed by the FDA

BIOMET 3i Endosseous Dental Implant.

Where it was distributed

Worldwide Distribution - USA and countries of United Arab Emirates, Argentina, Austria, Australia, Belgium, Brazil, Canada, Switzerland, Chile, Colombia, Costa Rica, Germany, Dominican Republic, Spain, France, Great Britain (UK), Greece, Hong Kong, Ireland, Israel, India, Italy, Japan, Korea (South), Lebanon, Liechtenstein, Lithuania, Mexico, Netherlands, Poland, Portugal, Paraguay, Saudi Arabia, Sweden, Singapore, Turkey, Taiwan, Uruguay, Vietnam and South Africa.

Questions about this recall

Why was it recalled?

Due to residual machining fluid identified by discoloration (darker in appearance) on the external dental implant surface.

Which lots are affected?

Model Number(s): Affected devices catalog names are: BNES310, BNET3210, BNPS6511, BNPS6513, BNPS6585, BNPT6510, BNPT6511, BNPT6513, BNPT6585, BNSS310, BNSS311, BNSS313, BNSS610, BNSS613, BNSS615, BNSS685, BNST3210, BNST3211, BNST3213, BNST3215, BNST3285, BNST611, BNST613, BNST685, BOPS6510, BOPS6511, BOPS6515, BOPS6585, BOPT6510, BOPT6511, BOPT6513, BOPT6585, BOSS310, BOSS311, BOSS313, BOSS315, BOSS611, BOSS613, BOSS685, BOST3210, BOST3211, BOST3213, BOST3215, BOST3285, BOST610, BOST611, BOST613, BOST685, FNT3210, FNT3211, IFNT3210, IFNT3211, IFNT3213, IFNT3215, IFNT3285, IFNT610, IFNT611, IFNT613, IFNT685, IFOSS5611, IFOS5685, IFOS610, IFOS611, IFOS615, IFOSM310, IFOSM311, IFOSM313, IFOSM385, IIOS5610, IIOS5611, INT3210, INT3211, INT3213, INT3215, INT3285, INT610, INT611, INT613, INT685, IOSM311, IOSM313, IOSM315, IOSM385, IOSS610, IOSS613, NIIITP6510, NIIITP6511, NIIITP6513, NIIITP6515, NIIITP6585, NINT3210, NINT3211, NINT3213, NINT3215, NINT610, NINT611, NINT613, NINT685, NIOSM311, NIOSM313, NIOSS610, NT3210, NT3211, OSM310, XIFNT3210, XIFNT3211, XIFNT3213, XIFNT3215, XIFNT3285, XIFN610, XIFNT611, XIFNT613, XIFNT685, XIFOSM310, XIFOSM311, XIFOSM313, XIFOSS610, XIFOSS611, XIIOS6510, XIIOS6511, XIIOS6513, XIITP6510, XIITP6511, XIITP6513. Manufactured dates from November 5, 2012 through August 21, 2013.

Where was it sold?

Worldwide Distribution - USA and countries of United Arab Emirates, Argentina, Austria, Australia, Belgium, Brazil, Canada, Switzerland, Chile, Colombia, Costa Rica, Germany, Dominican Republic, Spain, France, Great Britain (UK), Greece, Hong Kong, Ireland, Israel, India, Italy, Japan, Korea (South), Lebanon, Liechtenstein, Lithuania, Mexico, Netherlands, Poland, Portugal, Paraguay, Saudi Arabia, Sweden, Singapore, Turkey, Taiwan, Uruguay, Vietnam and South Africa.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Biomet 3i

All 222

Other Dental Devices recalls

All 426