CLASS II Device recall · Reported 16 Oct 2024 · FDA Enforcement Report

The Epione device v1.0.2 is a user controlled recalled

Quantum Surgical SAS is recalling The Epione device v1.0.2 is a user controlled, stereotactic accessory device intended to, distributed nationwide in the US. The recall was initiated on 5 Aug 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI/DI 37603054002NGXL, serial numbers: 0321015 and 0321017
Quantity recalled2 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0037-2025
FDA event ID
95194
Recalling firm
Quantum Surgical SAS, montpellier
Classification
Class II
Status
Completed
Recall initiated
5 Aug 2024
FDA classified
9 Oct 2024
Reported
16 Oct 2024
Type
Voluntary: Firm initiated
Customers notified by
Visit
Reason category
Other Reasons

Timeline

5 Aug 2024Recall initiated by company
9 Oct 2024Classified by FDA+65 days
16 Oct 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Quantum Surgical has become aware that a component (central axis) inside the Needle Guide product (product number 02-0027) is the subject of rust.

Product as listed by the FDA

The EPIONE device v1.0.2 is a user controlled, stereotactic accessory device intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures. During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.

Where it was distributed

US Nationwide distribution in the state of FL.

Nationwide Florida

Questions about this recall

Why was it recalled?

Quantum Surgical has become aware that a component (central axis) inside the Needle Guide product (product number 02-0027) is the subject of rust.

Which lots are affected?

UDI/DI 37603054002NGXL, serial numbers: 0321015 and 0321017

Where was it sold?

US Nationwide distribution in the state of FL.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Quantum Surgical SAS

All 2

Other CPAP & Ventilators recalls

All 554