CLASS II Device recall · Reported 27 Nov 2013 · FDA Enforcement Report
EPIQ 7 Ultrasound System recalled by Philips Medical Systems North
Philips Medical Systems North America is recalling EPIQ 7 Ultrasound System, distributed nationwide in the US. The recall was initiated on 17 Oct 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0368-2014
- FDA event ID
- 66662
- Recalling firm
- Philips Medical Systems North America, Bothell, WA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 17 Oct 2013
- FDA classified
- 20 Nov 2013
- Reported
- 27 Nov 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Imaging Equipment
Timeline
| 17 Oct 2013 | Recall initiated by company | |
| 20 Nov 2013 | Classified by FDA | +34 days |
| 27 Nov 2013 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A problem has been detected with Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. Using the International System Date Format may result in calculation errors in Gestational Age Gestational Age (GA) and Established Due Date (EDD).
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
EPIQ 7 Ultrasound System, Model 795200, 795201, 795202. Intended for diagnostic ultrasound imaging and fluid flow analysis.
Where it was distributed
Worldwide Distribution-USA (nationwide) including the states of CT, HI, IL, IN, KY, MA, ME, MI, NC, OR, PA, TX, and WA, and the countries of Australia, Austria, Belgium, Chile, Denmark, Finland, France, Germany, India, Ireland, Republic of Korea, Latvia, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.
Nationwide ConnecticutHawaiiIllinoisIndianaKentuckyMassachusettsMaineMichiganNorth CarolinaOregonPennsylvaniaTexasWashington
Questions about this recall
Is EPIQ 7 Ultrasound System recalled?
Yes. Philips Medical Systems North America recalled EPIQ 7 Ultrasound System on 17 Oct 2013. The recall is Class II and its status is Terminated. Recall number Z-0368-2014.
Why was it recalled?
A problem has been detected with Philips Model EPIQ 7 Ultrasound System, software version 1.0 and 1.0.1. Using the International System Date Format may result in calculation errors in Gestational Age Gestational Age (GA) and Established Due Date (EDD).
Which lots are affected?
TRACE ITEM SERIAL NUMBERS in the UNITED STATES: US713B0171, US713B0207, US813B0017, US813B0018, US813B0019, US813B0020, US813B0022, US813B0023, US813B0024, US813B0025, US813B0122, US813B0123, US813B0125, US813B0127, US813B0128, US813B0129, US813B0131, US813B0132, US813B0134, US813B0135, US813B0233, US813B0234, US813B0235, US813B0236, US813B0237, US813B0238, US813B0239, US813B0240, US813B0423, US813B0424, US813B0425, US813B0426, US813B0427, US813B0428, US813B0429, US813B0430, US813B0431, US813B0432, US813B0433, US813B0434, US913B0033, US913B0034, US913B0035, US913B0036, US913B0037, US913B0038, US913B0039, US913B0040, US913B0057, US913B0060, US913B0061, US913B0062, US913B0063, US913B0064, US913B0065, US913B0066, US913B0106, US913B0107, US913B0108, US913B0109, US913B0110, US913B0111, US913B0112, US913B0113, US913B0144, US913B0176, US913B0201, US913B0256, US913B0257, US913B0350, US913B0363, US913B0364, US913B0365, US913B0366, US913B0367, US913B0368, US913B0369, US913B0380, US913B0384, US913B0395, US913B0396, US913B0397, US913B0398, US913B0399, US913B0400, US913B0401, US913B0402, US913B0409, US913B0441, US913B0444, US913B0445, US913B0446, US913B0447, US913B0503, US913B0522, US913B0614, US913B0615, US913B0616, US913B0618, US913B0619, and USO13B0043. TRACE ITEM SERIAL NUMBERS INTERNATIONAL: US513B0224, US513B0295, US513B0387, US513B0560, US613B0299, US713B0085, US713B0086, US713B0118, US713B0119, US713B0120, US713B0122, US713B0123, US713B0126, US713B0149, US713B0171, US713B0172, US713B0192, US713B0193, US713B0194, US713B0195, US713B0204, US713B0205, US713B0206, US713B0208, US713B0324, US713B0325, US713B0358, US713B0359, US713B0437, US713B0439, US713B0440, US713B0532, US713B0537, US713B0573, US713B0574, US713B0575, US713B0576, US713B0577, US713B0578, US713B0579, US713B0580, US713B0581, US713B0582, US713B0583, US713B0584, US713B0586, US713B0587, US713B0588, US713B0589, US713B0590, US713B0592, US713B0593, US713B0594, US713B0595, US813B0015, US813B0016, US813B0032, US813B0037, US813B0038, US813B0039, US813B0121, US813B0151, US813B0152, US813B0153, US813B0154, US813B0155, US813B0156, US813B0157, US813B0158, US813B0159, US813B0160, US813B0161, US813B0162, US813B0163, US813B0164, US813B0183, US813B0184, US813B0185, US813B0186, US813B0187, US813B0188, US813B0189, US813B0190, US813B0191, US813B0192, US813B0193, US813B0194, US813B0195, US813B0196, US813B0232, US813B0396, US813B0397, US813B0398, US813B0399, US813B0400, US813B0401, US813B0402, US813B0403, US813B0404, US813B0405, US813B0406, US813B0407, US813B0408, US813B0409, US813B0410, US813B0411, US813B0468, US813B0469, US813B0470, US813B0471, US813B0472, US813B0473, US813B0474, US813B0475, US813B0476, US813B0477, US813B0478, US813B0479, US813B0480, US913B0026, US913B0027, US913B0028, US913B0029, US913B0030, US913B0031, US913B0032, US913B0041, US913B0051, US913B0059, US913B0085, US913B0086, US913B0101, US913B0102, US913B0103, US913B0104, US913B0105, US913B0130, US913B0132, US913B0133, US913B0134, US913B0135, US913B0137, US913B0138, US913B0139, US913B0140, US913B0141, US913B0142, US913B0143, US913B0177, US913B0178, US913B0179, US913B0185, US913B0202, US913B0216, US913B0217, US913B0218, US913B0219, US913B0220, US913B0221, US913B0222, US913B0223, US913B0232, US913B0252, US913B0253, US913B0254, US913B0255, US913B0259, US913B0260, US913B0299, US913B0300, US913B0301, US913B0302, US913B0303, US913B0316, US913B0317, US913B0318, US913B0319, US913B0332, US913B0333, US913B0336, US913B0337, US913B0338, US913B0339, US913B0340, US913B0349, US913B0359, US913B0360, US913B0385, US913B0408, US913B0458, US913B0459, US913B0486, US913B0487, US913B0488, US913B0489, US913B0490, US913B0499, US913B0504, US913B0505, US913B0506, US913B0565, US913B0566, US913B0567, US913B0568, USO13B0040, and USO13B0041.
Where was it sold?
Worldwide Distribution-USA (nationwide) including the states of CT, HI, IL, IN, KY, MA, ME, MI, NC, OR, PA, TX, and WA, and the countries of Australia, Austria, Belgium, Chile, Denmark, Finland, France, Germany, India, Ireland, Republic of Korea, Latvia, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, and United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips Medical Systems North America
All 5| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Philips HeartStart Home and OnSite recalled | Philips Medical Systems North America | Other | Nationwide |
| Device | CLASS II | Philips Model Epiq 5 Ultrasound System. Intended for diagnostic recalled | Philips Medical Systems North America | Other | Nationwide |
| Device | CLASS II | Philips Model Epiq 7 Ultrasound System. Intended for diagnostic recalled | Philips Medical Systems North America | Other | Nationwide |
| Device | CLASS II | Philips HeartStart FR3 Automated External Defibrillator (AED) recalled | Philips Medical Systems North America | Device Malfunction | 20 states |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device is recalled | Canon Medical System, USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederland B.v | Lead & Heavy Metals |