CLASS II ONGOING Device recall · Reported 30 Jul 2025 · FDA Enforcement Report
Estrone RIA recalled by Beckman Coulter
Beckman Coulter is recalling Estrone RIA, distributed in North Carolina. The recall was initiated on 12 Jun 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2152-2025
- FDA event ID
- 97132
- Recalling firm
- Beckman Coulter, Brea, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 12 Jun 2025
- FDA classified
- 21 Jul 2025
- Reported
- 30 Jul 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Other Medical Devices
Timeline
| 12 Jun 2025 | Recall initiated by company | |
| 21 Jul 2025 | Classified by FDA | +39 days |
| 30 Jul 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
Estrone RIA, REF: DSL8700. Estrone RIA is an IVD used for the quantitative measurement of estrone in human serum and plasma.
Where it was distributed
US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands
Questions about this recall
Is Estrone RIA recalled?
Yes. Beckman Coulter recalled Estrone RIA on 12 Jun 2025. The recall is Class II and its status is Ongoing. Recall number Z-2152-2025.
Why was it recalled?
A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
Which lots are affected?
UDI: 15099590211615/Lot: 250106D, 250203C, 250203D, 250303C, 250331C, 250428C, 250526C
Where was it sold?
US: North Carolina OUS: Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, Netherlands
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Beckman Coulter
All 334| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US) recalled | Beckman Coulter | Other | Nationwide |
| Device | CLASS II | REF # D23354, MicroScan LabPro V5.03 Panel Update 11 DVD Kit (US) on recalled | Beckman Coulter | Other | Nationwide |
| Device | CLASS II | Access Total T4 Calibrator recalled by Beckman Coulter | Beckman Coulter | Other | Nationwide |
Other Other Medical Devices recalls
All 13,771| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: Blood Draw Kit recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ortho Aspiration Kit recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep DYKA1054F recalled | Medline Industries | Lead & Heavy Metals | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Arterial Line Insertion Kit recalled | Medline Industries | Sterility | Nationwide |