CLASS II ONGOING Device recall · Reported 1 Apr 2026 · FDA Enforcement Report
Everest Bipolar Cutting Forceps recalled by Olympus Corporation of the
Olympus Corporation of the Americas is recalling Everest Bipolar Cutting Forceps, distributed nationwide in the US. The recall was initiated on 25 Feb 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1639-2026
- FDA event ID
- 98454
- Recalling firm
- Olympus Corporation of the Americas, Center Valley, PA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 25 Feb 2026
- FDA classified
- 25 Mar 2026
- Reported
- 1 Apr 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Heart & Vascular Devices
Timeline
| 25 Feb 2026 | Recall initiated by company | |
| 25 Mar 2026 | Classified by FDA | +28 days |
| 1 Apr 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Brand Name: Olympus Everest Bipolar Cutting Forceps w/Cord, 5mm 33 cm Product Name: Everest Bipolar Cutting Forceps Model/Catalog Number: 3006 Product Description: The OLYMPUS Everest Bipolar Cutting Forceps are intended to be passed through a 5 mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the bipolar outputs of compatible generators.
Where it was distributed
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
Questions about this recall
Why was it recalled?
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
Which lots are affected?
Model: 3006; UDI: 00821925035898; Lot Numbers: All unexpired;
Where was it sold?
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
6 other products · FDA event 98454This product is part of a recall event; the main page for the event is Olympus PKS Cutting Forceps PKS Cutting Forceps recalled.
More recalls from Olympus Corporation of the Americas
All 423| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Olympus M3 Gold Autoclavable Lateral Telescope recalled | Olympus Corporation of the Americas | Sterility | PA |
| Device | CLASS II | Olympus Tripp-Lite P580-025 25ft DisplayPort Cable recalled | Olympus Corporation of the Americas | Other | Nationwide |
| Device | CLASS II | Olympus M3 Gold Autoclavable Foroblique Telescope recalled | Olympus Corporation of the Americas | Sterility | PA |
| Device | CLASS II | Olympus BiCOAG Hemostasis Probe Fixed Pin Connector recalled | Olympus Corporation of the Americas | Other | Nationwide |
| Device | CLASS II | Olympus BiCOAG Hemostasis Probe Coaxial Connector recalled | Olympus Corporation of the Americas | Other | Nationwide |
Other Heart & Vascular Devices recalls
All 6,220| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TufTex¿ Over-the-Wire Embolectomy Catheter recalled | LeMaitre Vascular | Other | Nationwide |
| Device | CLASS II | Endovascular Pack kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Bard Ep Xt Decapolar Steerable, Diagnostic Ep Catheter, 6f recalled | Stryker Sustainability Solutions | Particulate Matter | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |