CLASS II ONGOING Device recall · Reported 1 Apr 2026 · FDA Enforcement Report

HALO PKS Cutting Forceps recalled by Olympus Corporation of the

Olympus Corporation of the Americas is recalling HALO PKS Cutting Forceps, distributed nationwide in the US. The recall was initiated on 25 Feb 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel: HACF0533; UDI: 00821925036390; Lot Numbers: All unexpired;
Quantity recalled2648 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1641-2026
FDA event ID
98454
Recalling firm
Olympus Corporation of the Americas, Center Valley, PA
Classification
Class II
Status
Ongoing
Recall initiated
25 Feb 2026
FDA classified
25 Mar 2026
Reported
1 Apr 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

25 Feb 2026Recall initiated by company
25 Mar 2026Classified by FDA+28 days
1 Apr 2026Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: Olympus HALO PKS Cutting Forceps, 5mm 33cm Product Name: HALO PKS Cutting Forceps Model/Catalog Number: HACF0533 Product Description: The HALO PKS Cutting Forceps are intended to be passed through a 5 mm laparoscopic cannula. Coagulation is achieved using electrosurgical energy under laparoscopic visualization. The device is intended to be used with the Gyrus ACMI G400 Workstation only.

Where it was distributed

Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.

Nationwide

Questions about this recall

Why was it recalled?

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Which lots are affected?

Model: HACF0533; UDI: 00821925036390; Lot Numbers: All unexpired;

Where was it sold?

Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

6 other products · FDA event 98454

This product is part of a recall event; the main page for the event is Olympus PKS Cutting Forceps PKS Cutting Forceps recalled.

More recalls from Olympus Corporation of the Americas

All 415

Other Heart & Vascular Devices recalls

All 6,199