CLASS II Device recall · Reported 2 Nov 2016 · FDA Enforcement Report

FemTouch delivery system includes the FemTouch Kit recalled by Lumenis

Lumenis is recalling FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip. The recall was initiated on 25 Sep 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAll serial numbers
Quantity recalled69 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0301-2017
FDA event ID
75384
Recalling firm
Lumenis, Yokneam
Classification
Class II
Status
Terminated
Recall initiated
25 Sep 2016
FDA classified
23 Oct 2016
Reported
2 Nov 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

25 Sep 2016Recall initiated by company
23 Oct 2016Classified by FDA+28 days
2 Nov 2016Published in enforcement report+10 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The cleaning instructions provided in FemTouch Operator's Manual might be ineffective for a fully assembled device and may lead to a potential for cross-contamination.

Product as listed by the FDA

FemTouch delivery system includes the FemTouch Kit (AC-1007415) and FemTouch" Tip (SPSA-20002590) Provide fractional treatments in gynecology.

Where it was distributed

Worldwide Distribution -- USA, Italy, Russian Fed., Australia, United Arab Emirates, Spain, United Kingdom, India, Kuwait, and Belarus.

Questions about this recall

Why was it recalled?

The cleaning instructions provided in FemTouch Operator's Manual might be ineffective for a fully assembled device and may lead to a potential for cross-contamination.

Which lots are affected?

All serial numbers

Where was it sold?

Worldwide Distribution -- USA, Italy, Russian Fed., Australia, United Arab Emirates, Spain, United Kingdom, India, Kuwait, and Belarus.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Lumenis

All 10
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIOtoLase Starter Kit recalled by LumenisLumenisSterilityPA, WI
DeviceCLASS IILumenis Pulse 120H 30 amp recalledLumenisOtherPA
DeviceCLASS IIPL Coupling Gel, I L (AX1009013) of (AX1009018) recalled by LumenisLumenisOtherNationwide
DeviceCLASS IIM22 System recalled by LumenisLumenisOther12 states
DeviceCLASS IIVersaCut+ Tissue Morcellator GA-0007500 The Lumenis recalledLumenisOther5 states

Other Other Medical Devices recalls

All 13,908