CLASS II ONGOING Device recall · Reported 12 Mar 2025 · FDA Enforcement Report

Lumenis Pulse 120H 30 amp recalled

Lumenis is recalling Lumenis Pulse 120H 30 amp, distributed in Pennsylvania. The recall was initiated on 23 Jan 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware version 2.4.1.0. (1) REF GA-2009996 - Serial number 1791, UDI-DI 07290109145518; (2) REF GA-0006802 - Serial number 1803, UDI-DI 07290109145525; and (3) REF-GA0008700 - Serial number 1818, UDI-DI 07290109145464.
Quantity recalled3 devices

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1290-2025
FDA event ID
96197
Recalling firm
Lumenis, Yokne'Am Ilit
Classification
Class II
Status
Ongoing
Recall initiated
23 Jan 2025
FDA classified
5 Mar 2025
Reported
12 Mar 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

23 Jan 2025Recall initiated by company
5 Mar 2025Classified by FDA+41 days
12 Mar 2025Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.

Product as listed by the FDA

(1) Lumenis Pulse 120H 30 amp, REF GA 2009996, Laser System; (2) Lumenis Pulse 120H, Moses 2.0 (DOM), REF GA0006802, Laser System; and (3) Lumenis Pulse 120H, REF GA-0008700, Laser System.

Where it was distributed

Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.

Pennsylvania

Questions about this recall

Why was it recalled?

The initial current in certain chargers may lead to overheating and damage to the power resistors on the Alternating Current (AC) controller during the power-on sequence of the laser system, resulting in the laser system to not power on.

Which lots are affected?

Software version 2.4.1.0. (1) REF GA-2009996 - Serial number 1791, UDI-DI 07290109145518; (2) REF GA-0006802 - Serial number 1803, UDI-DI 07290109145525; and (3) REF-GA0008700 - Serial number 1818, UDI-DI 07290109145464.

Where was it sold?

Distribution was made to PA. There was no government/military distribution. Foreign distribution was made to Mexico and Spain.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Lumenis

All 10
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIOtoLase Starter Kit recalled by LumenisLumenisSterilityPA, WI
DeviceCLASS IIPL Coupling Gel, I L (AX1009013) of (AX1009018) recalled by LumenisLumenisOtherNationwide
DeviceCLASS IIFemTouch delivery system includes the FemTouch Kit recalled by LumenisLumenisOther
DeviceCLASS IIM22 System recalled by LumenisLumenisOther12 states
DeviceCLASS IIVersaCut+ Tissue Morcellator GA-0007500 The Lumenis recalledLumenisOther5 states

Other Other Medical Devices recalls

All 13,908