CLASS II ONGOING Device recall · Reported 19 Jan 2022 · FDA Enforcement Report

Flat Panel Detector recalled by Canon Medical System, USA

Canon Medical System USA is recalling Flat Panel Detector, distributed nationwide in the US. The recall was initiated on 20 Dec 2021 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSID: 30039059, Alphenix System Serial Number: BGC20Y2010, & Flat Panel Detector Serial Number: C1B20Y2215 SID: 30030612, Alphenix System Serial Number: B8D20Z2001, & Flat Panel Detector Serial Number: C1B20Z2218; SID: 30030612, Alphenix System Serial Number: B8D20Z2001, & Flat Panel Detector Serial Number: C1B20Z2219; SID: 30046247, Alphenix System Serial Number: B1B20Z2001, & Flat Panel Detector Serial Number: C1B20Z2221; SID: 30064041, Alphenix System Serial Number: BGD2172011, & Flat Panel Detector Serial Number: C1B2172234; SID: 30065145, Alphenix System Serial Number: B1E2172002, & Flat Panel Detector Serial Number: C1B2172235; SID: 30058149, Alphenix System Serial Number: B1E2152001, & Flat Panel Detector Serial Number: C1B2152230; SID: 30002089, Alphenix System Serial Number: BHC20Z2003, & Flat Panel Detector Serial Number: C1B20Z2217.
Quantity recalled8

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0492-2022
FDA event ID
89218
Recalling firm
Canon Medical System USA, Tustin, CA
Classification
Class II
Status
Ongoing
Recall initiated
20 Dec 2021
FDA classified
12 Jan 2022
Reported
19 Jan 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

20 Dec 2021Recall initiated by company
12 Jan 2022Classified by FDA+23 days
19 Jan 2022Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Flat Panel Detector, Model Number: TFP-800A, used with the Alphenix Systems, Item Numbers: INFX-8000V, INFX-8000C, INFX-8000F, and INFX-8000H.

Where it was distributed

US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.

Nationwide FloridaLouisianaMontanaNew JerseyOhioWest Virginia

Questions about this recall

Why was it recalled?

Abnormal images such as horizontal stripes may be displayed and recovery may not not be possible by restarting the system. The worst case scenario is an operator who is working without a workable image may remove a catheter from the patient's blood vessel and damage the blood vessel.

Which lots are affected?

SID: 30039059, Alphenix System Serial Number: BGC20Y2010, & Flat Panel Detector Serial Number: C1B20Y2215 SID: 30030612, Alphenix System Serial Number: B8D20Z2001, & Flat Panel Detector Serial Number: C1B20Z2218; SID: 30030612, Alphenix System Serial Number: B8D20Z2001, & Flat Panel Detector Serial Number: C1B20Z2219; SID: 30046247, Alphenix System Serial Number: B1B20Z2001, & Flat Panel Detector Serial Number: C1B20Z2221; SID: 30064041, Alphenix System Serial Number: BGD2172011, & Flat Panel Detector Serial Number: C1B2172234; SID: 30065145, Alphenix System Serial Number: B1E2172002, & Flat Panel Detector Serial Number: C1B2172235; SID: 30058149, Alphenix System Serial Number: B1E2152001, & Flat Panel Detector Serial Number: C1B2152230; SID: 30002089, Alphenix System Serial Number: BHC20Z2003, & Flat Panel Detector Serial Number: C1B20Z2217.

Where was it sold?

US Nationwide distribution in the states of DE, FL, LA, MT, NJ, OH, and WV.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Canon Medical System USA

All 51

Other Other Medical Devices recalls

All 13,908