CLASS II Device recall · Reported 8 Aug 2018 · FDA Enforcement Report

FlowGate2 Balloon Guide Catheter, UPN recalled by Stryker Neurovascular

Stryker Neurovascular is recalling FlowGate2 Balloon Guide Catheter, UPN 90485 (8Fx85cm size) and 90495 (8Fx95cm size), distributed nationwide in the US. The recall was initiated on 11 Apr 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLots 1352, 1630, 1631, 1802, 1970, 2652, 2653, 2654, 2655, 2656, 2657, 2658, 2659, 3169, 3940, 3941
Quantity recalled968 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2569-2018
FDA event ID
80353
Recalling firm
Stryker Neurovascular, Fremont, CA
Classification
Class II
Status
Terminated
Recall initiated
11 Apr 2018
FDA classified
28 Jul 2018
Reported
8 Aug 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

11 Apr 2018Recall initiated by company
28 Jul 2018Classified by FDA+108 days
8 Aug 2018Published in enforcement report+11 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

There may be product manufactured with the outer diameter of the balloon catheter exceeding the defined specification at a customer location. If the user experiences resistance between the FlowGate and an interfacing device, the device would be removed and a second device would be used, which could potentially prolong a procedure.

Product as listed by the FDA

FlowGate2 Balloon Guide Catheter, UPN 90485 (8Fx85cm size) and 90495 (8Fx95cm size)

Where it was distributed

Nationwide

Nationwide

Questions about this recall

Why was it recalled?

There may be product manufactured with the outer diameter of the balloon catheter exceeding the defined specification at a customer location. If the user experiences resistance between the FlowGate and an interfacing device, the device would be removed and a second device would be used, which could potentially prolong a procedure.

Which lots are affected?

Lots 1352, 1630, 1631, 1802, 1970, 2652, 2653, 2654, 2655, 2656, 2657, 2658, 2659, 3169, 3940, 3941

Where was it sold?

Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Neurovascular

All 109

Other Heart & Vascular Devices recalls

All 6,199