CLASS II Device recall · Reported 8 Aug 2018 · FDA Enforcement Report
FlowGate2 Balloon Guide Catheter, UPN recalled by Stryker Neurovascular
Stryker Neurovascular is recalling FlowGate2 Balloon Guide Catheter, UPN 90485 (8Fx85cm size) and 90495 (8Fx95cm size), distributed nationwide in the US. The recall was initiated on 11 Apr 2018 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2569-2018
- FDA event ID
- 80353
- Recalling firm
- Stryker Neurovascular, Fremont, CA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 11 Apr 2018
- FDA classified
- 28 Jul 2018
- Reported
- 8 Aug 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Heart & Vascular Devices
Timeline
| 11 Apr 2018 | Recall initiated by company | |
| 28 Jul 2018 | Classified by FDA | +108 days |
| 8 Aug 2018 | Published in enforcement report | +11 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
There may be product manufactured with the outer diameter of the balloon catheter exceeding the defined specification at a customer location. If the user experiences resistance between the FlowGate and an interfacing device, the device would be removed and a second device would be used, which could potentially prolong a procedure.
Product as listed by the FDA
FlowGate2 Balloon Guide Catheter, UPN 90485 (8Fx85cm size) and 90495 (8Fx95cm size)
Where it was distributed
Nationwide
Questions about this recall
Why was it recalled?
There may be product manufactured with the outer diameter of the balloon catheter exceeding the defined specification at a customer location. If the user experiences resistance between the FlowGate and an interfacing device, the device would be removed and a second device would be used, which could potentially prolong a procedure.
Which lots are affected?
Lots 1352, 1630, 1631, 1802, 1970, 2652, 2653, 2654, 2655, 2656, 2657, 2658, 2659, 3169, 3940, 3941
Where was it sold?
Nationwide
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Stryker Neurovascular
All 109| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Inzone Detachment System recalled by Stryker Neurovascular | Stryker Neurovascular | Device Malfunction | Nationwide |
| Device | CLASS II | The Trevo Trak 21 Microcatheter is a single-lumen recalled | Stryker Neurovascular | Packaging Defects | Nationwide |
| Device | CLASS II | The Synchro Neuro Guidewires are a steerable guidewire recalled | Stryker Neurovascular | Other | Nationwide |
| Device | CLASS II | The Synchro Select Guidewires are a steerable guidewire recalled | Stryker Neurovascular | Other | Nationwide |
| Device | CLASS II | The Synchro2 Guidewires are a steerable guidewire family recalled | Stryker Neurovascular | Other | Nationwide |
Other Heart & Vascular Devices recalls
All 6,199| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Getinge PVC Thoracic Catheter, Soft PVC Right Angle recalled | Atrium Medical | Potency & Assay | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes: *Bardex¿ I.C recalled | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Products that contain the Pericare Wipes recalled by C.R. Bard | C.R. Bard | Other | Nationwide |
| Device | CLASS II | Getinge PVC Thoracic Catheter, Firm PVC Straight recalled | Atrium Medical | Potency & Assay | Nationwide |