CLASS I Device recall · Reported 26 Mar 2014 · FDA Enforcement Report

FreeStyle Blood Glucose Monitors System recalled by Abbott Diabetes Care

Abbott Diabetes Care is recalling FreeStyle Blood Glucose Monitors System, distributed nationwide in the US. The recall was initiated on 19 Feb 2014 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesPart number CAT11001-0; all serial numbers
Quantity recallednot provided. Device has not been shipped in the last 5 years.

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1211-2014
FDA event ID
67472
Recalling firm
Abbott Diabetes Care, Alameda, CA
Classification
Class I
Status
Terminated
Recall initiated
19 Feb 2014
FDA classified
18 Mar 2014
Reported
26 Mar 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Labeling Errors

Timeline

19 Feb 2014Recall initiated by company
18 Mar 2014Classified by FDA+27 days
26 Mar 2014Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.

Product as listed by the FDA

FreeStyle Blood Glucose Monitors System. For in vitro diagnostic use only. Store at room temperature below 86 degrees F (30 degrees C). Use test strips only within the system operating temperature range as outlined in your Owner's Booklet. Meter made in China; Distributed by Abbott Diabetes Care, Alameda, CA The FreeStyle mete should only be used with FreeStyle Test Strips and FreeStyle Control Solution. Using other brands of test strips and control solutions with the FreeStyle Mete can product inaccurate results. The FreeStyle Glucose meter is incorporated into the OmniPod insulin Management system (manufactured by Insulet Corporation, Bedford, MA and is intended for subcutaneous delivery of insulin. Intended to monitor blood glucose from samples taken from the body. IVD use only,

Where it was distributed

Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.

Nationwide

Questions about this recall

Is FreeStyle Blood Glucose Monitors System recalled?

Yes. Abbott Diabetes Care recalled FreeStyle Blood Glucose Monitors System on 19 Feb 2014. The recall is Class I and its status is Terminated. Recall number Z-1211-2014.

Why was it recalled?

Abbott Diabetes Care has identified through internal testing and investigation that all non-applied voltage legacy meters have the potential to produce out of range control solution results and/or erroneously low blood glucose results when used in conjunction with Free Style test strips lots within expiry.

Which lots are affected?

Part number CAT11001-0; all serial numbers

Where was it sold?

Worldwide Distribution: US (nationwide) and countries of: Austria, Canada, Germany, Israel, Netherlands, Norway, Sweden, Switzerland, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.