CLASS II ONGOING Device recall · Reported 18 Nov 2026 · FDA Enforcement Report
Tandem Mobi Insulin Pump with Interoperable Technology recalled
Tandem Diabetes Care is recalling Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile, distributed nationwide in the US. The recall was initiated on 13 Oct 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0437-2027
- FDA event ID
- 100091
- Recalling firm
- Tandem Diabetes Care, San Diego, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 13 Oct 2026
- FDA classified
- 12 Nov 2026
- Reported
- 18 Nov 2026
- Reason category
- Device Malfunction
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 13 Oct 2026 | Recall initiated by company | |
| 12 Nov 2026 | Classified by FDA | +30 days |
| 18 Nov 2026 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred ( Later ) portion is delivered, but the total insulin delivered may be less than intended, and the application does not provide a distinct alert that the immediate portion was omitted, potentially resulting in unrecognized under-delivery of insulin and hyperglycemia.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Tandem Mobi Insulin Pump with Interoperable Technology (with Android Tandem Mobi Mobile app) Model/Catalog Number: 1015472
Where it was distributed
Worldwide - US Nationwide and the country of Czech Republic.
Questions about this recall
Why was it recalled?
A software defect in the Tandem Mobi mobile application for Android (versions 3.5.4, 3.5.2.2, and 3.5.2.1) in which the immediate ( Now ) portion of an Extended Bolus is incorrectly reduced to 0 units when the entered glucose value equals the user s active target glucose value; the deferred ( Later ) portion is delivered, but the total insulin delivered may be less than intended, and the application does not provide a distinct alert that the immediate portion was omitted, potentially resulting in unrecognized under-delivery of insulin and hyperglycemia.
Which lots are affected?
Lot Code: UDI-DI: 00389152547244 Software versions: 3.5.4, 3.5.2.2, and 3.5.2.1
Where was it sold?
Worldwide - US Nationwide and the country of Czech Republic.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 25 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Tandem Diabetes Care
All 15| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Tandem Mobi Insulin Pump recalled by Tandem Diabetes Care | Tandem Diabetes Care | Device Malfunction | Nationwide |
| Device | CLASS II | Tandem t:slim mobile application t:slim Mobile Application, iOS recalled | Tandem Diabetes Care | Device Malfunction | Nationwide |
| Device | CLASS II | Tandem pumps are battery-operated infusion pumps capable recalled | Tandem Diabetes Care | Device Malfunction | |
| Device | CLASS I | Tandem Mobi Insulin Pump: Description recalled by Tandem Diabetes Care | Tandem Diabetes Care | Device Malfunction | 52 states |
| Device | CLASS II | t:slim X2 insulin pump with Interoperable Technology recalled | Tandem Diabetes Care | Device Malfunction | Nationwide |
Other Insulin Pumps & Glucose Monitors recalls
All 470| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Tandem Mobi Insulin Pump recalled by Tandem Diabetes Care | Tandem Diabetes Care | Device Malfunction | Nationwide |
| Device | CLASS II | The MiniMed Go app for iOS or MiniMed Go app recalled | Medtronic MiniMed | Device Malfunction | Nationwide |
| Device | CLASS II | MiniMed Flex Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Labeling Errors | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG Strip recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | ANNOUNCED | StatStrip Glucose Hospital Meter System recalled by Nova BiomedicalNova Biomedical StatStrip | Nova Biomedical | Device Malfunction |