CLASS II ONGOING Device recall · Reported 1 Jul 2026 · FDA Enforcement Report
G7 Continuous Glucose Monitoring System Dexcom G7 US Retail recalled
Dexcom is recalling Dexcom G7 US Retail Sensor/Transmitter. The recall was initiated on 27 May 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2574-2026
- FDA event ID
- 98968
- Recalling firm
- Dexcom, San Diego, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 27 May 2026
- FDA classified
- 24 Jun 2026
- Reported
- 1 Jul 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 27 May 2026 | Recall initiated by company | |
| 24 Jun 2026 | Classified by FDA | +28 days |
| 1 Jul 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring System
Where it was distributed
Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom.
Questions about this recall
Is G7 Continuous Glucose Monitoring System Dexcom G7 US Retail Sensor/Transmitter recalled?
Yes. Dexcom recalled G7 Continuous Glucose Monitoring System Dexcom G7 US Retail Sensor/Transmitter on 27 May 2026. The recall is Class II and its status is Ongoing. Recall number Z-2574-2026.
Why was it recalled?
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.
Which lots are affected?
Lot Code: UDI-DI: 00386270004109 Lot Numbers: 1725204004 and 1725069002
Where was it sold?
Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Dexcom
All 29| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | ONE+ iOS CGM App recalled by Dexcom | Dexcom | Device Malfunction | Nationwide |
| Device | CLASS II | G7 Continuous Glucose Monitoring System Dexcom G7 CGM iOS App recalled | Dexcom | Device Malfunction | Nationwide |
| Device | CLASS II | G7 Continuous Glucose Monitoring System Dexcom G7 CGM watchOS App recalled | Dexcom | Device Malfunction | Nationwide |
| Device | CLASS I | G6 Continuous Glucose Monitoring System Dexcom G6 recalled | Dexcom | Device Malfunction | Nationwide |
| Device | CLASS I | G7 Continuous Glucose Monitoring System Dexcom G7 CGM iOS App recalled | Dexcom | Lead & Heavy Metals | Nationwide |
Other Insulin Pumps & Glucose Monitors recalls
All 459| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Lead & Heavy Metals | Nationwide |
| Device | CLASS I | Omnipod (Eros) Pod recalled by Insulet | Insulet | Lead & Heavy Metals | Nationwide |