CLASS II ONGOING Device recall · Reported 20 May 2026 · FDA Enforcement Report
ONE+ iOS CGM App recalled by Dexcom
Dexcom is recalling Dexcom ONE+ iOS CGM App, distributed nationwide in the US. The recall was initiated on 14 Apr 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2165-2026
- FDA event ID
- 98639
- Recalling firm
- Dexcom, San Diego, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 14 Apr 2026
- FDA classified
- 11 May 2026
- Reported
- 20 May 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 14 Apr 2026 | Recall initiated by company | |
| 11 May 2026 | Classified by FDA | +27 days |
| 20 May 2026 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Brand Name: Dexcom ONE+ Continuous Glucose Monitoring System Product Name: Dexcom ONE+ iOS CGM App Model/Catalog Number: SW14244 Software Version: versions 1.5.0 and 1.6.0 Product Description: Dexcom ONE+ iOS CGM App Component: Dexcom ONE+ Continuous Glucose Monitoring System
Where it was distributed
SW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2: Worldwide - US Nationwide including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Austria, Australia, Belgium, Bahrain, Canada, Czech Republic, Germany, Denmark, Estonia, Finland, France, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, Kuwait, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom. SW14245 Dexcom ONE iOS App versions 1.5.0 and 1.6.0 - Only International distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Czech Republic, Estonia, France, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Malaysia, Monaco, Netherlands, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Spain, Sweden, United Kingdom.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Is ONE+ iOS CGM App recalled?
Yes. Dexcom recalled ONE+ iOS CGM App on 14 Apr 2026. The recall is Class II and its status is Ongoing. Recall number Z-2165-2026.
Why was it recalled?
A software defect in which a low-priority file I/O check blocks the higher-priority thread responsible for processing estimated glucose values (EGVs) and trend arrows. When the app is running in the background, EGV notifications may queue up and then "replay" sequentially when the user brings the app to the foreground, causing the Glucose Compass to flicker through previous EGVs. In more severe instances, the app may perceive signal loss and deploy signal loss mitigations. This delay in EGV and alert processing creates a risk that users may miss detection of a hypoglycemic or hyperglycemic event, or make treatment decisions based on outdated glucose information.
Which lots are affected?
Lot Code: SW14244 versions 1.5.0 and 1.6.0 - UDI-DI: 00386270004307
Where was it sold?
SW12300 & SW13355 Dexcom G7 iOS and watchOS app versions 2.9.0 through 2.11.2: Worldwide - US Nationwide including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Austria, Australia, Belgium, Bahrain, Canada, Czech Republic, Germany, Denmark, Estonia, Finland, France, Hong Kong, Ireland, Israel, Iceland, Italy, Jordan, Japan, Kuwait, Luxembourg, Malta, Netherlands, Norway, New Zealand, Oman, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Slovenia, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom. SW14245 Dexcom ONE iOS App versions 1.5.0 and 1.6.0 - Only International distribution to the countries of Andorra, Australia, Belgium, Bulgaria, Czech Republic, Estonia, France, Greece, Croatia, Ireland, Italy, Lithuania, Latvia, Malaysia, Monaco, Netherlands, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Spain, Sweden, United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Dexcom
All 29| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | G7 Continuous Glucose Monitoring System Dexcom G7 US Retail recalled | Dexcom | Sterility | CA |
| Device | CLASS II | G7 Continuous Glucose Monitoring System Dexcom G7 CGM iOS App recalled | Dexcom | Device Malfunction | Nationwide |
| Device | CLASS II | G7 Continuous Glucose Monitoring System Dexcom G7 CGM watchOS App recalled | Dexcom | Device Malfunction | Nationwide |
| Device | CLASS I | G6 Continuous Glucose Monitoring System Dexcom G6 recalled | Dexcom | Device Malfunction | Nationwide |
| Device | CLASS I | G7 Continuous Glucose Monitoring System Dexcom G7 CGM iOS App recalled | Dexcom | Lead & Heavy Metals | Nationwide |
Other Insulin Pumps & Glucose Monitors recalls
All 459| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Other | Nationwide |
| Device | CLASS II | MiniMed Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Lead & Heavy Metals | Nationwide |
| Device | CLASS I | Omnipod (Eros) Pod recalled by Insulet | Insulet | Lead & Heavy Metals | Nationwide |