CLASS I ONGOING Device recall · Reported 5 Nov 2025 · FDA Enforcement Report

G6 Continuous Glucose Monitoring System Dexcom G6 recalled

Dexcom is recalling G6 Continuous Glucose Monitoring System Dexcom G6 and G6 Pro Android US CGM App, distributed nationwide in the US. The recall was initiated on 28 Aug 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesLot Code: UDI: SW11678 - 00386270000811
Quantity recalled333

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0318-2026
FDA event ID
97607
Recalling firm
Dexcom, San Diego, CA
Classification
Class I
Status
Ongoing
Recall initiated
28 Aug 2025
FDA classified
30 Oct 2025
Reported
5 Nov 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

28 Aug 2025Recall initiated by company
30 Oct 2025Classified by FDA+63 days
5 Nov 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System

Where it was distributed

This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.

Nationwide

Questions about this recall

Why was it recalled?

A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).

Which lots are affected?

Lot Code: UDI: SW11678 - 00386270000811

Where was it sold?

This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Dexcom

All 29

Other Insulin Pumps & Glucose Monitors recalls

All 466