CLASS II ONGOING Device recall · Reported 1 Jan 2025 · FDA Enforcement Report

Galaxy System (GAL-001) recalled by Noah Medical

Noah Medical is recalling Galaxy System (GAL-001), distributed in 15 states. The recall was initiated on 28 Oct 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1506, CRB-1508, CRB-1516, CRB-1517, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1527, CRB-1528, CRB-1529, CRB-1530, CRB-1532,
Quantity recalled18

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0742-2025
FDA event ID
95585
Recalling firm
Noah Medical, San Carlos, CA
Classification
Class II
Status
Ongoing
Recall initiated
28 Oct 2024
FDA classified
20 Dec 2024
Reported
1 Jan 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

28 Oct 2024Recall initiated by company
20 Dec 2024Classified by FDA+53 days
1 Jan 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Galaxy System (GAL-001). Intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.

Where it was distributed

US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC

CaliforniaConnecticutIdahoIndianaMassachusettsMarylandMissouriNorth CarolinaNew JerseyNew YorkOhioPennsylvaniaSouth CarolinaTennesseeWest Virginia

Questions about this recall

Why was it recalled?

Bronchoscopic visualization system power supplies have defective semiconductor components that may fail unexpectedly, which could result in an aborted procedure since it is a non recoverable failure. Manual removal of the bronchoscope when the system is not functioning can increase the chance of injury to the tissues of the lung as the scope is withdrawn including the possibility of pneumothorax.

Which lots are affected?

UDI-DI: 00850048825048. Serial Numbers: CRB-1502, CRB-1503, CRB-1506, CRB-1508, CRB-1516, CRB-1517, CRB-1519, CRB-1520, CRB-1521, CRB-1522, CRB-1523, CRB-1524, CRB-1525, CRB-1527, CRB-1528, CRB-1529, CRB-1530, CRB-1532,

Where was it sold?

US: SC, OH, PA, TN, MO, IN, CA, WV, NY, MD, MA, NJ, ID, CT, NC

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Noah Medical

All 5
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIGalaxy System recalled by Noah MedicalNoah MedicalOther20 states
DeviceCLASS IIGalaxy System (GAL-001) recalled by Noah MedicalNoah MedicalDevice Malfunction16 states
DeviceCLASS IIGalaxy bronchoscope (Product Galb-001) used recalled by Noah MedicalNoah MedicalOtherNationwide
DeviceCLASS IIGalaxy System recalled by Noah MedicalNoah MedicalOtherNationwide

Other Test Kits & Diagnostics recalls

All 2,661