CLASS I Device recall · Reported 5 Mar 2014 · FDA Enforcement Report

Giraffe and Panda Bag and Mask Resuscitation System;Giraffe recalled

GE Healthcare is recalling Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant, distributed nationwide in the US. The recall was initiated on 1 Oct 2013 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesQABS50026, QABR50066, QABR50068, QABR50133, QABS50023, QABR50102, QABR50103, QABR50064, QABR50065, QABR50067, QABS50070, QABS50069, QABR50275, QABS50027, QABR50274
Quantity recalled12, 373 units total, 8,807 = US, 3,394 = OUS, 172 = unknown

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1071-2014
FDA event ID
67004
Recalling firm
GE Healthcare, Waukesha, WI
Classification
Class I
Status
Terminated
Recall initiated
1 Oct 2013
FDA classified
25 Feb 2014
Reported
5 Mar 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

1 Oct 2013Recall initiated by company
25 Feb 2014Classified by FDA+147 days
5 Mar 2014Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

Product as listed by the FDA

GE Healthcare - Giraffe and Panda Bag and Mask Resuscitation System;Giraffe Stand-alone Infant Resuscitation System. Product Usage: The Bag and Mask Resuscitation System provides the basic equipment required for pulmonary resuscitation of infants. Pulmonary resuscitation includes practices necessary to establish a clear airway and provide oxygen or air/oxygen mixtures and/or manual ventilation to the infant. These are clinical practices that represent the established standard of care.

Where it was distributed

Worldwide Distribution - USA Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL,GA. HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and the countries of Albania, Austria, Australia, Bahrain, Belarus, Belgium, Canada, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Greece, Guatemala, Germany, Honduras, Hong Kong, Hungry, India, Indonesia, Ireland, Italy, Kazakhstan, Korea, Lithuania, Malaysia, Mexico, Netherlands, Norway, Peru, Poland, Romania, Russia, Saudi Arabia, South Africia, Spain, Sweden, Switzerland Turkey, United Arab, United Kingdom, and Venezuela.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMaine

Questions about this recall

Why was it recalled?

Oxygen and air wall inlet fittings on the back panel of the resuscitation systems were reversed during assembly.

Which lots are affected?

QABS50026, QABR50066, QABR50068, QABR50133, QABS50023, QABR50102, QABR50103, QABR50064, QABR50065, QABR50067, QABS50070, QABS50069, QABR50275, QABS50027, QABR50274

Where was it sold?

Worldwide Distribution - USA Nationwide including the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL,GA. HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; and the countries of Albania, Austria, Australia, Bahrain, Belarus, Belgium, Canada, China, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Greece, Guatemala, Germany, Honduras, Hong Kong, Hungry, India, Indonesia, Ireland, Italy, Kazakhstan, Korea, Lithuania, Malaysia, Mexico, Netherlands, Norway, Peru, Poland, Romania, Russia, Saudi Arabia, South Africia, Spain, Sweden, Switzerland Turkey, United Arab, United Kingdom, and Venezuela.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 67004
ReportedClassRecallCompanyReasonWhere
CLASS IGiraffe Warmer;Panda Warmer;Panda iRes Warmer recalled by GE HealthcareGE HealthcareOtherNationwide

More recalls from GE Healthcare

All 512

Other Masks, Gloves & PPE recalls

All 447