CLASS I ONGOING Device recall · Reported 31 Aug 2022 · FDA Enforcement Report

Hamilton-C6 Intensive Care Ventilator recalled by Hamilton Medical AG

Hamilton Medical is recalling Hamilton-C6 Intensive Care Ventilator, distributed nationwide in the US. The recall was initiated on 27 Jun 2022 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesUDI-DI: 07630002808590, Serial Numbers: 5663, 6995, 6907, 6952, 7177, 7772, 6932, 6157, 8658, 7307, 5939, 6858, 6231, 8620, 6349, 8356, 6480, 8322, 8306, 9100, 9101, 9099, 9098, 6219, 7894, 7861, 8003, 6723, 7802, 4833, 7265, 7980, 7666, 7382, 6774, 7358, 7719, 6896, 6925, 9073, 9062, 9298, 9294, 9301, 9292, 9293, 9290, 9300, 9299, 9291, 9231, 9132, 9108, 8310, 8604, 8255, 8599, 7156, 8313, 8344, 8590, 8087, 8348, 8490, 8151, 7174, 8594, 8477, 8122, 8063, 9353, 9230, 9329, 9243, 9235, 9236, 9337, 9228, 9224, 9229, 9225, 9103, 8413, 6149, 8178, 8401, 7827, 8406, 8409, 7826, 6416, 7831, 6280, 9128, 9260, 9257, 9125, 9120, 9259, 9258, 9354, 8325, 8297, 7777, 6212, 6516, 6871, 8099, 6562, 8258, 5679, 5681, 7808, 7115, 8601, 7349, 7130, 6658, 7581, 7819, 8346, 6537, 8235, 7410, 6868, 8049, 8281, 7966, 7787, 9241, 9242, 9054, 9075, 9131, 9121, 9137, 9134, 9133, 9118, 6350, 7122, 7117, 6790, 7614, 8121, 7884, 8102, 8720, 8715, 8726, 8718, 8719, 9248, 9244, 9127, 9138, 9126, 8256, 8371, 9107, 9109, 9036, 9110, 9069, 9063, 9067, 8097, 9129, 9130, 9459, 8723, 8722, 8713, 8702, 8725, 8714, 8727, 8706, 8517, 8717, 7822, 8148, 7695, 8261, 7165, 6944, 7420, 6937, 7150, 6953, 7201, 6965, 7004, 7196, 8849, 8597, 8850, 9106, 9192, 9239, 9237, 9233, 8311, 7898, 7893, 6641, 6505, 6075, 7986, 8005, 7888, 7887, 9535, 9579, 9508, 9576, 9553, 9577, 9578, 9545, 9582, 9590, 9575, 7845, 7740, 8082, 8070, 6671, 7608, 6566, 8535, 8616, 7821, 6579, 7586, 7994, 7066, 7138, 9102, 9105, 14784, 8603, 8846, 8844, 8847, 8845, 8115, 7113, 7809, 7938, 7993, 7252, 8568, 7754, 7033, 8052, 6829, 8249, 9077, 9080, 9076, 9051, 9064, 9081, 9050, 9071, 9091, 7836, 7656, 9124, 9462, 9297, 9325, 9303, 8866, 9348, 9352, 9324, 9227, 6093, 6159, 6274, 6276, 6309, 6320, 6327, 6464, 6512, 6545, 6674, 6789, 7136, 7173, 7191, 7273, 7279, 7422, 7432, 7569, 7572, 7587, 7644, 7676, 7782, 7783, 7784, 7786, 7788, 7789, 7790, 7792, 7793, 7794, 7795, 7796, 7797, 7798, 7799, 7801, 7806, 7807, 7812, 7813, 7837, 7901, 7923, 7969, 8182, 8343, 8394, 8395, 8399, 8420, 8421, 8422, 8423, 8424, 8497, 8519, 8593, 8661, 8668, 8670, 8673, 8675, 8676, 8678, 8707, 8709, 8712, 9028, 9029, 9032, 9033, 9042, 9043, 9044, 9046
Quantity recalled358

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1525-2022
FDA event ID
90576
Recalling firm
Hamilton Medical, Bonaduz
Classification
Class I
Status
Ongoing
Recall initiated
27 Jun 2022
FDA classified
19 Aug 2022
Reported
31 Aug 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

27 Jun 2022Recall initiated by company
19 Aug 2022Classified by FDA+53 days
31 Aug 2022Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

HAMILTON-C6 Intensive Care Ventilator, REF: 160021

Where it was distributed

US nationwide distribution including Puerto Rico.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed

Which lots are affected?

UDI-DI: 07630002808590, Serial Numbers: 5663, 6995, 6907, 6952, 7177, 7772, 6932, 6157, 8658, 7307, 5939, 6858, 6231, 8620, 6349, 8356, 6480, 8322, 8306, 9100, 9101, 9099, 9098, 6219, 7894, 7861, 8003, 6723, 7802, 4833, 7265, 7980, 7666, 7382, 6774, 7358, 7719, 6896, 6925, 9073, 9062, 9298, 9294, 9301, 9292, 9293, 9290, 9300, 9299, 9291, 9231, 9132, 9108, 8310, 8604, 8255, 8599, 7156, 8313, 8344, 8590, 8087, 8348, 8490, 8151, 7174, 8594, 8477, 8122, 8063, 9353, 9230, 9329, 9243, 9235, 9236, 9337, 9228, 9224, 9229, 9225, 9103, 8413, 6149, 8178, 8401, 7827, 8406, 8409, 7826, 6416, 7831, 6280, 9128, 9260, 9257, 9125, 9120, 9259, 9258, 9354, 8325, 8297, 7777, 6212, 6516, 6871, 8099, 6562, 8258, 5679, 5681, 7808, 7115, 8601, 7349, 7130, 6658, 7581, 7819, 8346, 6537, 8235, 7410, 6868, 8049, 8281, 7966, 7787, 9241, 9242, 9054, 9075, 9131, 9121, 9137, 9134, 9133, 9118, 6350, 7122, 7117, 6790, 7614, 8121, 7884, 8102, 8720, 8715, 8726, 8718, 8719, 9248, 9244, 9127, 9138, 9126, 8256, 8371, 9107, 9109, 9036, 9110, 9069, 9063, 9067, 8097, 9129, 9130, 9459, 8723, 8722, 8713, 8702, 8725, 8714, 8727, 8706, 8517, 8717, 7822, 8148, 7695, 8261, 7165, 6944, 7420, 6937, 7150, 6953, 7201, 6965, 7004, 7196, 8849, 8597, 8850, 9106, 9192, 9239, 9237, 9233, 8311, 7898, 7893, 6641, 6505, 6075, 7986, 8005, 7888, 7887, 9535, 9579, 9508, 9576, 9553, 9577, 9578, 9545, 9582, 9590, 9575, 7845, 7740, 8082, 8070, 6671, 7608, 6566, 8535, 8616, 7821, 6579, 7586, 7994, 7066, 7138, 9102, 9105, 14784, 8603, 8846, 8844, 8847, 8845, 8115, 7113, 7809, 7938, 7993, 7252, 8568, 7754, 7033, 8052, 6829, 8249, 9077, 9080, 9076, 9051, 9064, 9081, 9050, 9071, 9091, 7836, 7656, 9124, 9462, 9297, 9325, 9303, 8866, 9348, 9352, 9324, 9227, 6093, 6159, 6274, 6276, 6309, 6320, 6327, 6464, 6512, 6545, 6674, 6789, 7136, 7173, 7191, 7273, 7279, 7422, 7432, 7569, 7572, 7587, 7644, 7676, 7782, 7783, 7784, 7786, 7788, 7789, 7790, 7792, 7793, 7794, 7795, 7796, 7797, 7798, 7799, 7801, 7806, 7807, 7812, 7813, 7837, 7901, 7923, 7969, 8182, 8343, 8394, 8395, 8399, 8420, 8421, 8422, 8423, 8424, 8497, 8519, 8593, 8661, 8668, 8670, 8673, 8675, 8676, 8678, 8707, 8709, 8712, 9028, 9029, 9032, 9033, 9042, 9043, 9044, 9046

Where was it sold?

US nationwide distribution including Puerto Rico.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Hamilton Medical

All 34

Other CPAP & Ventilators recalls

All 568