CLASS II ONGOING Device recall · Reported 3 May 2023 · FDA Enforcement Report
Hamilton-C6, REF recalled by Hamilton Medical AG
Hamilton Medical is recalling Hamilton-C6, REF: 160021, distributed nationwide in the US. The recall was initiated on 14 Mar 2023 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1429-2023
- FDA event ID
- 91925
- Recalling firm
- Hamilton Medical, Bonaduz, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 14 Mar 2023
- FDA classified
- 25 Apr 2023
- Reported
- 3 May 2023
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 14 Mar 2023 | Recall initiated by company | |
| 25 Apr 2023 | Classified by FDA | +42 days |
| 3 May 2023 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
HAMILTON-C6, REF: 160021
Where it was distributed
US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.
Nationwide CaliforniaFloridaGeorgiaIdahoKansasMassachusettsMissouriNorth CarolinaNew MexicoNevadaNew YorkPennsylvaniaSouth CarolinaVirginiaWyoming
Questions about this recall
Why was it recalled?
Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.
Which lots are affected?
Software Version: 1.2.1, UDI-DI: 07630002808590
Where was it sold?
US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Hamilton Medical
All 34| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | IntelliCuff, Portable automatic cuff pressure controller recalled | Hamilton Medical | Packaging Defects | Nationwide |
| Device | CLASS I | Breathing circuit set, coaxial recalled by Hamilton Medical AG | Hamilton Medical | Device Malfunction | Nationwide |
| Device | CLASS I | Hamilton C6 recalled by Hamilton Medical AG | Hamilton Medical | Device Malfunction | Nationwide |
| Device | CLASS I | Breathing circuit set, coaxial, 1.80 m, with flow sensor recalled | Hamilton Medical | Device Malfunction | Nationwide |
| Device | CLASS I | Ventilator Hamilton-C6, PN recalled by Hamilton Medical AG | Hamilton Medical | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |