CLASS II ONGOING Device recall · Reported 3 May 2023 · FDA Enforcement Report

Hamilton-C6, REF recalled by Hamilton Medical AG

Hamilton Medical is recalling Hamilton-C6, REF: 160021, distributed nationwide in the US. The recall was initiated on 14 Mar 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware Version: 1.2.1, UDI-DI: 07630002808590
Quantity recalled103

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1429-2023
FDA event ID
91925
Recalling firm
Hamilton Medical, Bonaduz, N/A
Classification
Class II
Status
Ongoing
Recall initiated
14 Mar 2023
FDA classified
25 Apr 2023
Reported
3 May 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

14 Mar 2023Recall initiated by company
25 Apr 2023Classified by FDA+42 days
3 May 2023Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

HAMILTON-C6, REF: 160021

Where it was distributed

US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.

Nationwide CaliforniaFloridaGeorgiaIdahoKansasMassachusettsMissouriNorth CarolinaNew MexicoNevadaNew YorkPennsylvaniaSouth CarolinaVirginiaWyoming

Questions about this recall

Why was it recalled?

Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.

Which lots are affected?

Software Version: 1.2.1, UDI-DI: 07630002808590

Where was it sold?

US Nationwide distribution in the states of CA, MA, ID, NC, GA, FL, NV, PA, MO, KS, SC, NY, NM, WY, VA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.