CLASS I ONGOING Device recall · Reported 27 Sep 2023 · FDA Enforcement Report

Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator recalled

Hamilton Medical is recalling Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator, distributed nationwide in the US. The recall was initiated on 26 Jul 2023 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesUDI: 07630002801850 & 07630002813532 / Affected Serial Numbers: 15929 17600 17661 17677 17698 17699 17741 17850 17877 17913 17925 17955 17957 17969 17975 17976 17982 17985 17986 17991 17993 17995 18009 18166 18183 18189 18191 18199 18224 18244 18251 18255 18256 18260 18263 18272 18273 18276 18288 18289 18305 18308 18333 18340 18349 18360 18363 18377 18378 18383 18414 18415 18419 18441 18451 18458 18461 18502 18537 18544 18550 18778 18784 19237 19290 19291 19292 19293 19294 19298 19304 19305 19470 19483 19490 26117
Quantity recalled76 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2526-2023
FDA event ID
92873
Recalling firm
Hamilton Medical, Reno, NV
Classification
Class I
Status
Ongoing
Recall initiated
26 Jul 2023
FDA classified
20 Sep 2023
Reported
27 Sep 2023
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Lead & Heavy Metals

Timeline

26 Jul 2023Recall initiated by company
20 Sep 2023Classified by FDA+56 days
27 Sep 2023Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Lead and other heavy metals accumulate in the body. Children are most at risk; even low levels can affect development.

Product as listed by the FDA

Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator

Where it was distributed

Distribution US nationwide, including Puerto Rico. OUS: Not provided

Nationwide Puerto Rico

Questions about this recall

Is Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator recalled?

Yes. Hamilton Medical recalled Hamilton-T1, REF: 161006, Intensive Care and Transport Ventilator on 26 Jul 2023. The recall is Class I and its status is Ongoing. Recall number Z-2526-2023.

Why was it recalled?

Degrading capacitors on the control board of ventilators and spare parts may leak electrolyte onto the control board causing a short circuit on the board and/or the capacitor to lose function and lead to interruption of ventilation.

Which lots are affected?

UDI: 07630002801850 & 07630002813532 / Affected Serial Numbers: 15929 17600 17661 17677 17698 17699 17741 17850 17877 17913 17925 17955 17957 17969 17975 17976 17982 17985 17986 17991 17993 17995 18009 18166 18183 18189 18191 18199 18224 18244 18251 18255 18256 18260 18263 18272 18273 18276 18288 18289 18305 18308 18333 18340 18349 18360 18363 18377 18378 18383 18414 18415 18419 18441 18451 18458 18461 18502 18537 18544 18550 18778 18784 19237 19290 19291 19292 19293 19294 19298 19304 19305 19470 19483 19490 26117

Where was it sold?

Distribution US nationwide, including Puerto Rico. OUS: Not provided

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.