CLASS II ONGOING Device recall · Reported 27 Jul 2022 · FDA Enforcement Report

Hill-Rom PRO+ 36" MRS Surface With X-Ray mattress recalled

Baxter Healthcare is recalling Hill-Rom PRO+ 36" MRS Surface With X-Ray mattress, distributed nationwide in the US. The recall was initiated on 24 Jun 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesREF P7924A01, UDI 00887761977907; Serial numbers: W218BP4042, W218BP4054, W218BP4056, W218BP4059, W218BP4060, W218BP4061, W218BP4058, W218BP4048, W218BP4069, W218BP4053, W218BP4041, W218BP4043, W218BP4044, W218BP4045, W218BP4046, W218BP4049, W218BP4051, W218BP4052, W218BP4055, W218BP4057, W218BP4062, W218BP4063, W218BP4065, W218BP4066, W218BP4067, W336BP6549, W336BP6550, W336BP6555, W336BP6556, W336BP6557, W336BP6558, W336BP6560, W336BP6561, W336BP6562, W336BP6563, W218BP4050, W336BP6548, W337BP6609, W336BP6541, W336BP6546, W336BP6540, W336BP6542, W336BP6543, and W336BP6544. REF P7924ARENT01, UDI 00887761976863; Serial numbers: W223BP4081, W223BP4084, W242BP4397, W242BP4398, W242BP4401, W242BP4403, W242BP4405, W242BP4407, W242BP4408, W242BP4409, W242BP4410, W231BP4233, W231BP4234, W231BP4241, W231BP4243, W231BP4244, W231BP4248, W231BP4249, W231BP4252, W231BP4253, W235BP4259, W235BP4260, W235BP4265, W235BP4266, W242BP4381, W242BP4387, W242BP4391, W242BP4392, W242BP4393, W242BP4394, W242BP4395, W242BP4396, W242BP4402, W242BP4404, W243BP4412, W243BP4413, W243BP4414, W243BP4415, W243BP4416, W231BP4213, W231BP4224, W231BP4228, W231BP4229, W231BP4230, W231BP4242, W231BP4245, W231BP4246, W231BP4247, W231BP4250, W231BP4251, W242BP4390, W243BP4418, W243BP4419, W243BP4420, W243BP4421, W243BP4428, W223BP4079, W224BP4102, W224BP4108, W231BP4202, W231BP4203, W231BP4204, W231BP4207, W231BP4209, W231BP4217, W231BP4218, W231BP4220, W231BP4221, W231BP4222, W231BP4223, W231BP4227, W231BP4254, W231BP4255, W231BP4256, W223BP4091, W224BP4103, W231BP4210, W231BP4211, W231BP4214, W231BP4216, W231BP4219, W231BP4257, W235BP4261, W235BP4262, W235BP4263, W235BP4264, W239BP4361, W239BP4365, W239BP4367, W239BP4373, W239BP4375, W242BP4382, W242BP4383, W242BP4385, W242BP4386, W223BP4070, W223BP4071, W223BP4072, W223BP4073, W223BP4074, W223BP4077, W223BP4082, W223BP4083, W223BP4085, W223BP4086, W223BP4087, W223BP4088, W223BP4090, W223BP4092, W223BP4093, W224BP4096, W224BP4098, W224BP4099, W224BP4100, W224BP4101, W224BP4104, W224BP4105, W224BP4109, W224BP4110, W224BP4111, W231BP4205, W231BP4206, W231BP4208, W231BP4212, W231BP4215, W224BP4095, W224BP4097, W224BP4112, W224BP4113, W224BP4114, W224BP4106, W224BP4107, W239BP4364, W239BP4366, W239BP4369, W239BP4371, W239BP4372, W239BP4362, W239BP4368, W239BP4370, W239BP4360, W239BP4363, W239BP4374, W242BP4378, W242BP4379, W242BP4380, W242BP4384, W242BP4399, W242BP4400, W223BP4075, W223BP4080, W223BP4076, W223BP4078, W224BP4094, W224BP4115, and W224BP4116.
Quantity recalled200 mattresses

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1411-2022
FDA event ID
90511
Recalling firm
Baxter Healthcare, Deerfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
24 Jun 2022
FDA classified
20 Jul 2022
Reported
27 Jul 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

24 Jun 2022Recall initiated by company
20 Jul 2022Classified by FDA+26 days
27 Jul 2022Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Hill-Rom PRO+ 36" MRS Surface With X-Ray mattress, REF P7924A01; and REF P7924ARENT01, PRO+ MRS Rental Surface with X-Ray mattress.

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. There was government distribution but no military distribution. The country of Canada.

Nationwide ArizonaCaliforniaColoradoConnecticutGeorgiaIowaIndianaKentuckyLouisianaMarylandMichiganMinnesotaMissouriNorth CarolinaNew JerseyNew MexicoNew YorkOhioOklahomaRhode Island

Questions about this recall

Is Hill-Rom PRO+ 36" MRS Surface With X-Ray mattress recalled?

Yes. Baxter Healthcare recalled Hill-Rom PRO+ 36" MRS Surface With X-Ray mattress on 24 Jun 2022. The recall is Class II and its status is Ongoing. Recall number Z-1411-2022.

Why was it recalled?

The failure mode on the affected mattresses can cause a reduction in the performance of the Microclimate Management feature used to pull heat and moisture away from the patient resulting in an increased patient risk of pressure ulcer development.

Which lots are affected?

REF P7924A01, UDI 00887761977907; Serial numbers: W218BP4042, W218BP4054, W218BP4056, W218BP4059, W218BP4060, W218BP4061, W218BP4058, W218BP4048, W218BP4069, W218BP4053, W218BP4041, W218BP4043, W218BP4044, W218BP4045, W218BP4046, W218BP4049, W218BP4051, W218BP4052, W218BP4055, W218BP4057, W218BP4062, W218BP4063, W218BP4065, W218BP4066, W218BP4067, W336BP6549, W336BP6550, W336BP6555, W336BP6556, W336BP6557, W336BP6558, W336BP6560, W336BP6561, W336BP6562, W336BP6563, W218BP4050, W336BP6548, W337BP6609, W336BP6541, W336BP6546, W336BP6540, W336BP6542, W336BP6543, and W336BP6544. REF P7924ARENT01, UDI 00887761976863; Serial numbers: W223BP4081, W223BP4084, W242BP4397, W242BP4398, W242BP4401, W242BP4403, W242BP4405, W242BP4407, W242BP4408, W242BP4409, W242BP4410, W231BP4233, W231BP4234, W231BP4241, W231BP4243, W231BP4244, W231BP4248, W231BP4249, W231BP4252, W231BP4253, W235BP4259, W235BP4260, W235BP4265, W235BP4266, W242BP4381, W242BP4387, W242BP4391, W242BP4392, W242BP4393, W242BP4394, W242BP4395, W242BP4396, W242BP4402, W242BP4404, W243BP4412, W243BP4413, W243BP4414, W243BP4415, W243BP4416, W231BP4213, W231BP4224, W231BP4228, W231BP4229, W231BP4230, W231BP4242, W231BP4245, W231BP4246, W231BP4247, W231BP4250, W231BP4251, W242BP4390, W243BP4418, W243BP4419, W243BP4420, W243BP4421, W243BP4428, W223BP4079, W224BP4102, W224BP4108, W231BP4202, W231BP4203, W231BP4204, W231BP4207, W231BP4209, W231BP4217, W231BP4218, W231BP4220, W231BP4221, W231BP4222, W231BP4223, W231BP4227, W231BP4254, W231BP4255, W231BP4256, W223BP4091, W224BP4103, W231BP4210, W231BP4211, W231BP4214, W231BP4216, W231BP4219, W231BP4257, W235BP4261, W235BP4262, W235BP4263, W235BP4264, W239BP4361, W239BP4365, W239BP4367, W239BP4373, W239BP4375, W242BP4382, W242BP4383, W242BP4385, W242BP4386, W223BP4070, W223BP4071, W223BP4072, W223BP4073, W223BP4074, W223BP4077, W223BP4082, W223BP4083, W223BP4085, W223BP4086, W223BP4087, W223BP4088, W223BP4090, W223BP4092, W223BP4093, W224BP4096, W224BP4098, W224BP4099, W224BP4100, W224BP4101, W224BP4104, W224BP4105, W224BP4109, W224BP4110, W224BP4111, W231BP4205, W231BP4206, W231BP4208, W231BP4212, W231BP4215, W224BP4095, W224BP4097, W224BP4112, W224BP4113, W224BP4114, W224BP4106, W224BP4107, W239BP4364, W239BP4366, W239BP4369, W239BP4371, W239BP4372, W239BP4362, W239BP4368, W239BP4370, W239BP4360, W239BP4363, W239BP4374, W242BP4378, W242BP4379, W242BP4380, W242BP4384, W242BP4399, W242BP4400, W223BP4075, W223BP4080, W223BP4076, W223BP4078, W224BP4094, W224BP4115, and W224BP4116.

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, GA, IA, IN, KY, LA, MD, MI, MN, MO, NC, NJ, NM, NM, NY, OH, OK, RI, SD, TN, TX, VT, and WI. There was government distribution but no military distribution. The country of Canada.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.